ILX Lightwave MPS-8033 – Precision Fiber Optic Source
The ILX Lightwave MPS-8033 is a precision fiber optic source designed for demanding applications. It offers stable and accurate optical power, essential for testing and characterization of fiber optic components and systems.
- Model
- ILX Lightwave MPS-8033
- Calibration
- Available on request
- Price
- $1,450
- Status
- In stock
- RFQ ref
- MPS-8033
This unit
The ILX Lightwave MPS-8033 is a microprocessor-controlled precision fiber optic source engineered for component testing, power meter calibration, and laboratory characterization. It delivers stable, accurate optical power across multiple wavelength options with exceptional short-term and long-term power stability. The instrument supports both continuous wave and internally modulated operation, with remote programming via GPIB for seamless integration into automated test systems.
Technical Specifications
Optical Output- Wavelength options: WDM C-Band (1527.98–1564.26 nm), WDM L-Band (1564.27–1610.06 nm), 850 nm, 980 nm, 1310 nm, 1480 nm, 1550 nm, and custom wavelengths upon request
- Wavelength accuracy: ±20 nm (options /03–/06); ±0.10 nm (WDM options)
- Maximum output power: 20 mW (0 dBm, 10 dBm, or 13 dBm) for WDM and DFB lasers; 1 mW for 850/980/1480 nm FP lasers; +13 dBm (20 mW) for ASE models
- Power adjustment: 10 dB range
- Power stability: ±0.005 dB (15 min); ±0.030 dB (24 hours); ±0.03 dB (12 hours, ASE models)
- Thermal stability: 0.2 dB
- Output connector: FC/APC single-mode fiber Operating Modes & Modulation
- Continuous Wave (CW) and internal digital modulation
- Internal modulation frequencies: 270 Hz, 1 kHz, 2 kHz
- External modulation via GPIB: 160 Hz to 15 kHz
- Modulation duty cycle: 50% ±2%
- Edge jitter: 0.01% Display & Control
- Power display: milliwatts (mW), decibel-milliwatts (dBm), or relative dB
- Laser diode control: precision temperature and current electronics
- Warm-up time: 2 hours recommended
Key Features
- GPIB remote interface for full front-panel function access and instrument-specific command support
- Increased display resolution through remote programming
- TTL trigger output (ASE models) with 5 ns jitter and BNC connector
- Compact form factor: 88 × 212 × 269 mm; <4.7 kg
Typical Applications
- Optical component characterization and testing
- Power meter calibration
- Automated test system integration
- Laboratory fiber optic measurements
Compatibility & Integration GPIB interface enables direct host computer control and integration into production and laboratory test environments. All operating modes and settings accessible remotely. Power supply: 90–125 VAC, 50–60 Hz.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | 0 dBm, 10 dBm, 13 dBm (DFB/WDM options) |
| +13 dBm maximum (ASE models) | Voltage / Current Range | 50–60 Hz, 90–125 VAC | Wavelength Range: 1310 nm, 1480 nm, 1527.98–1564.26 nm (WDM C-Band), 1550 nm, 1564.27–1610.06 nm (WDM L-Band), 850 nm, 980 nm |
ILX Lightwave MPS-8033 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the ILX Lightwave MPS-8033. Calibration covers output power at specified wavelengths, wavelength accuracy, 15-minute output stability, and modulation frequency accuracy where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common ILX Lightwave MPS-8033 faults: laser diode end-of-life (output below spec), fiber output connector damage, output instability. Common repairs include laser diode replacement when output falls below spec, fiber output connector repair, and thermal management service.
ILX Lightwave MPS-8033 price
This ILX Lightwave MPS-8033 is listed at $1,450. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
ILX Lightwave MPS-8033 calibration cost
NIST-traceable calibration for the ILX Lightwave MPS-8033 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output power at specified wavelengths, wavelength accuracy, 15-minute output stability, and modulation frequency accuracy where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →