Advantest R3162 9kHz to 8GHz Spectrum Analyzer

Advantest Spectrum Analyzers. Frequency range: 9 kHz to 8 GHz, Accuracy: 1% or less, Accuracy: ±Span 1% + Resolution Bandwidth 15% + 60 Hz.

Model
Advantest R3162
Calibration
Available on request
Price
$3,779
RFQ ref
R3162
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Advantest R3162 is a 9 kHz to 8 GHz spectrum analyzer engineered for RF and microwave characterization across development, production, and service environments. Built on direct digital synthesis architecture, it delivers 1% frequency span accuracy and maintains a 117 dB dynamic range through exceptional distortion performance: -80 dBc second-order harmonic distortion, -80 dBc two-signal third-order intermodulation distortion, and a 0 dBm input 1 dB gain compression point.


Technical Specifications

Frequency Coverage & Accuracy

  • Frequency range: 9 kHz to 8 GHz

  • Frequency span accuracy: 1% or less

  • Frequency reading accuracy: ±Span 1% + Resolution Bandwidth 15% + 60 Hz

  • Signal purity: 105 dBc/Hz at 20 kHz offset; 100 dBc/Hz at 10 kHz offset Level Measurement

  • Measuring range: +30 dBm to average noise level

  • Total level accuracy: ±1.5 dB maximum

  • Calibration signal level accuracy: -20 dBm ±0.3 dB

  • Reference level range (preamplifier off): -64 dBm to +40 dBm logarithmic; 141.1 µV to 22.36 V linear

  • Reference level range (preamplifier on): -84.4 dBm to -20.4 dBm logarithmic; 13.47 mV to +22.35 mV linear

  • Average noise level (1 kHz RBW, preamplifier off): -118 dBm + 2f (GHz) dB Resolution & Bandwidth

  • Resolution bandwidth: 1 kHz to 3 MHz in 1-3-10 sequence; optional 30 Hz, 100 Hz, 300 Hz (3 dB); 200 Hz (6 dB)

  • Video bandwidth: 10 Hz to 3 MHz in 1-3-10 sequence Measurement Performance

  • Sweep time: 20 ms to 1000 s

  • Trace update rate: Up to 20 traces per second

  • Maximum input level (ATT 5 dB, preamplifier off): +30 dBm, 50 VDC max

  • Maximum input level (ATT 5 dB, preamplifier on): +13 dBm, 50 VDC max


Key Features



  • Real-time waveform observation at 20 updates per second

  • Selectable input attenuation to 50 dB with 0.3 dB switching error

  • Optional narrow resolution bandwidth modes for enhanced spectral selectivity

  • Auto-calibration reduces IF gain error to 0.5 dB and RBW switching level error to 0.5 dB

  • Logarithmic scale indication accuracy of 0.5 dB from 0 to -20 dB


Typical Applications



  • Component and subsystem RF characterization

  • Harmonic and intermodulation distortion measurement

  • Transmitter and receiver performance validation

  • EMI/RFI troubleshooting and compliance verification


Compatibility & Integration The R3162 accepts input signals up to +30 dBm with input ATT 5 dB (preamplifier off) and 50 VDC maximum, supporting integration into multi-instrument test systems.

Specifications

ParameterValue
Frequency Range9 kHz to 8 GHz | Power Range (dBm): '-84.4 dBm to +40 dBm (preamplifier off)
-118 dBm average noise floor | Voltage / Current Range50 VDC max input

Advantest R3162 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Advantest R3162. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Advantest R3162 price

This Advantest R3162 is listed at $3,779. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Advantest R3162 calibration cost

NIST-traceable calibration for the Advantest R3162 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →