IFR 2030 10kHz to 1.35GHz Signal Generator

IFR Signal Generators. Frequency Range: 10 kHz to 1.35 GHz, Resolution: 0.1 Hz, Resolution: 0.1 dB.

Model
IFR 2030
Calibration
Available on request
Price
$1,979
Status
In stock
RFQ ref
2030
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The IFR 2030 is a signal generator delivering precise RF signal generation from 10 kHz to 1.35 GHz with 0.1 Hz frequency resolution. It supports AM, FM, and phase modulation with flexible modulation oscillators operating from 0.1 Hz to 500 kHz. Output power spans -144 dBm to +13 dBm with 0.1 dB resolution, and the instrument withstands reverse power up to 50 W at VSWR 5:1. An RF carrier on/off switch provides complete output control. FM capabilities include wideband operation at 10 MHz bandwidth, FM deviation to 1 MHz (up to 21 MHz) or 1% of carrier frequency, and both AC and DC-coupled modes with offset correction. Phase modulation reaches 10 radians across a 10 kHz bandwidth. AM operation reduces maximum output to +7 dBm at full depth. The instrument supports single, dual, composite, and dual composite modulation modes. Optional modules enable pulse modulation, GSM/PCN (GMSK Bt 0.3) with 270.833 kHz clock rate, avionics waveforms for VOR/ILS testing, and DME generation. Non-volatile memory stores instrument configuration. GPIB (IEEE-488.2) interface enables automated test system integration and remote operation.

Technical Specifications

Frequency Range: 10 kHz to 1.35 GHz

Frequency Resolution: 0.1 Hz

Output Power: -144 dBm to +13 dBm

Output Power Resolution: 0.1 dB

Output Protection: 50 W reverse power capability at VSWR 5:1

Modulation Oscillator: 0.1 Hz to 500 kHz

FM Deviation: Up to 1 MHz (≤21 MHz); 1% of carrier frequency (>21 MHz)

FM Bandwidth: 10 MHz wideband; 300 kHz at 1 dB

Phase Modulation Bandwidth: 10 kHz; maximum 10 radians

AM Maximum Output: +7 dBm at full modulation depth

Key Features

AM, FM, and phase modulation with single/dual/composite modes

DC-coupled FM with offset correction

Dedicated RF carrier on/off switch

Non-volatile memory for settings storage

Optional second modulation oscillator

Optional pulse modulation (including slow rise time variant)

Optional GSM/PCN GMSK Bt 0.3 modulation with internal data generator

Optional avionics modulation for VOR/ILS system testing

Optional DME signal generation

Typical Applications

RF component characterization and receiver testing

Radio communications system validation

Instrument Landing System (ILS) and VHF Omni Range (VOR) beacon testing

GSM/PCN mobile communications testing

Distance Measuring Equipment (DME) validation

Compatibility & Integration

GPIB (IEEE-488.2) interface for automated test system integration

Remote operation and control via standard IEEE-488.2 commands

Specifications

ParameterValue
Frequency Range10 kHz to 1.35 GHz | Power Range (dBm): '-144 dBm to +13 dBm

IFR 2030 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the IFR 2030. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common IFR 2030 faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

IFR 2030 price

This IFR 2030 is listed at $1,979. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

IFR 2030 calibration cost

NIST-traceable calibration for the IFR 2030 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →