Agilent/Keysight 3468A Option 001

Agilent / Keysight Laboratory Measurement Instruments. Voltage: 30 mV to 300 V with 1 µV sensitivity. Input resis, Current: 300 µA to 3 A.

Model
Agilent / Keysight 3468A
Calibration
Available on request
Price
$750
Status
In stock
RFQ ref
3468A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent/Keysight 3468A Option 001 is a 5½ to 3½ digit digital multimeter delivering six measurement functions: DC voltage, true RMS AC voltage, 2-wire resistance, 4-wire resistance, DC current, and true RMS AC current. Display resolution is selectable across 5½, 4½, or 3½ digits, trading measurement speed against noise rejection as application demands. Autoranging operates across all functions.


The instrument features closed-box electronic calibration - eliminating internal adjustments and executable manually or remotely via the built-in HP-IL interface. Remote programmability provides control over functions, ranges, and measurement modifiers. Option 001 includes a rechargeable battery pack for field operation.


Technical Specifications

DC Voltage: 30 mV to 300 V with 1 µV sensitivity. Input resistance exceeds 10 GΩ on ranges to 3 V; 10 MΩ ±1% on 30 V and 300 V ranges. Maximum non-destructive input: 300 Vrms or 450 V peak. Accuracy specified as ±(0.0035% of reading + 2 counts) on the 3 V range over a 24-hour period. AC Voltage (True RMS): 0.3 V to 300 V, 1 µV sensitivity. Bandwidth extends 20 Hz to 100 kHz on all ranges; 20 Hz to 300 kHz on the 30 V range. Crest factor >4:1 at full scale. Input impedance: 1 MΩ ±1% in parallel with 10% of full scale. Resistance (2-wire/4-wire): 30 Ω to 30 MΩ with 1 mΩ sensitivity. Open circuit voltage <6.5 V. Test current varies: 1 mA (300 Ω, 3 kΩ), 1 µA (300 kΩ, 3 MΩ), 100 nA (30 MΩ). 2-wire accuracy matches 4-wire plus maximum 100 mΩ offset. DC Current: 300 µA to 3 A, 10 µA sensitivity. Maximum non-destructive input: 3 A from <250 V source; fuse protected. AC Current (True RMS): 300 µA to 3 A. General: Operates 0 to 55°C, 95% RH (0 to 40°C). AC line powered 48 to 440 Hz, 86 to 250 V. Dimensions: 98.4 mm H × 238.1 mm W × 276.2 mm D. Weight: 2.1 kg.

Key Features



  • Selectable digit resolution for speed/noise optimization

  • Closed-box calibration eliminates drift-prone adjustments

  • HP-IL interface for remote integration into test systems

  • Rechargeable battery option for portable environments

  • True RMS AC measurements to 300 kHz

  • 4-wire resistance measurement eliminates lead resistance error


Compatibility & Integration HP-IL connectivity enables integration into Hewlett-Packard automated test systems and compatible instrumentation networks.

Specifications

ParameterValue
Frequency Range20 Hz to 100 kHz (300 mV to 30 V AC ranges)
AC current bandwidth to 20 kHz | Voltage / Current Range3 A
AC Current300 mA, 3 A
Resistance300 Ω to 30 MΩ, DC Voltage: 300 mV to 300 V
AC Voltage300 mV to 300 V
DC Current300 mA

Agilent / Keysight 3468A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 3468A. Calibration covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / Keysight 3468A faults: blown input fuse, worn input jacks, display segment failure, relay contact wear on bench DMMs. Common faults include worn input jacks (banana socket contact wear on the mA and COM terminals), blown input fuses (especially the 440 mA HRC fuse), and display segment failures. Keithley 2000/2001/2002 bench DMMs commonly develop relay contact wear affecting resistance accuracy after extended service.

Agilent / Keysight 3468A price

This Agilent / Keysight 3468A is listed at $750. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight 3468A calibration cost

NIST-traceable calibration for the Agilent / Keysight 3468A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. Turnaround is 1 to 3 business days after receipt. See full calibration details →