HP / Agilent 4262A Digital LCR Meter
HP / Agilent Laboratory Measurement Instruments. Accuracy: 0.2% to 0.3% of reading.
- Model
- HP / Agilent 4262A
- Calibration
- Available on request
- Price
- $952
- Status
- In stock
- RFQ ref
- 4262A
This unit
The HP/Agilent 4262A is a digital LCR meter for precision measurement of inductance, capacitance, resistance, dissipation factor, and quality factor across component testing and production environments. Fully automatic ranging selects optimal measurement range and equivalent circuit mode - series or parallel - while front-panel offset adjustments compensate for stray capacitance (0 to 10 pF) and residual inductance (0 to 1 µH) of test fixtures. The instrument delivers 0.2% to 0.3% basic accuracy with a 3½ digit LED display and supports multiple measurement frequencies: 120 Hz (or 100 Hz with Option 010), 1 kHz, and 10 kHz at ±3% frequency accuracy.
Technical Specifications
- Measurement Parameters: Inductance (L), Capacitance (C), Resistance (R), Dissipation Factor (D), Quality Factor (Q)
- Equivalent Circuit Modes: Series and Parallel
- Measurement Ranges: Capacitance 0.01 pF to 19.99 mF; Inductance 0.01 µH to 1999 H; Resistance 1 mΩ to 19.99 MΩ
- Frequencies: 120 Hz (or 100 Hz with Option 010), 1 kHz, 10 kHz at ±3% accuracy
- Basic Accuracy: 0.2% to 0.3% of reading
- Test Signal Level: 1 V RMS (0.05 V RMS in Cp mode)
- Internal DC Bias: 1.5 V, 2.2 V, 6 VDC (±5%)
- External DC Bias: Up to +40 VDC maximum
- Measurement Speed: 1 kHz/10 kHz: 120–260 ms (C/L), 120–160 ms (R); 120/100 Hz: 900 ms (C/L), 700 ms (R)
- Ranging Time: 1 kHz/10 kHz: 180 ms/step; 120/100 Hz: 670 ms/step
- Reading Rate: ~30 ms internal trigger; external trigger available
- Warm-up Time: 15 minutes
- Temperature Coefficient: Accuracy doubles between 0–55°C relative to 23°C ±5°C
Key Features
- Automatic and manual ranging modes
- Internal self-test capability for functional verification
- Deviation measurements display difference between stored and measured values
- Offset adjustments for test fixture stray capacitance and residual inductance
Typical Applications Electrolytic and ceramic capacitor testing, filter coil measurement, pulse transformer characterization, dry cell internal resistance, semiconductor junction capacitance, and standard LCR component analysis.
Compatibility & Integration
- Option 001: BCD Data Output
- Option 004: Digital Comparator (GO/NO-GO testing)
- Option 010: 100 Hz lower test frequency
- Option 101: IEEE-488 GPIB Interface (incompatible with Options 001 and 004)
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 100 Hz to 10 kHz | Voltage / Current Range: '+40 VDC maximum external bias |
HP / Agilent 4262A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 4262A. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 4262A faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.
HP / Agilent 4262A price
This HP / Agilent 4262A is listed at $952. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 4262A calibration cost
NIST-traceable calibration for the HP / Agilent 4262A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →