Wilton 6647A-40 Sweep Generator, 10MHz to 20gHz
The Wiltron 6647A-40 is a sweep generator that provides precise and reliable frequency control from 10 MHz to 20 GHz. Ideal for testing and aligning microwave components and systems, it delivers stable and accurate signals for a variety of applications.
- Model
- Wiltron 6647A-40
- Calibration
- Available on request
- Price
- $4,480
- Status
- In stock
- RFQ ref
- 6647A-40
This unit
The Wiltron 6647A-40 is a programmable sweep generator delivering precise frequency control from 10 MHz to 18.6 GHz for microwave component testing and system alignment. This instrument combines hard-wired circuitry within a frequency module with flexible sweep modes, multiple marker types, and three leveling control options to meet demanding laboratory and production environments.
Technical Specifications
- Frequency Range: 10 MHz to 18.6 GHz
- Output Power: >10 dBm (internally leveled); >−82 dBm with Option 2 (70 dB Attenuator)
- Frequency Accuracy: ±10 MHz (CW Mode); ±15 MHz (Sweep Mode ≤50 MHz)
- Sweep Time: 0.01 to 99 seconds
- Power Level Accuracy: Within 1 dB
- Signal Purity: Harmonics typically −40 dBc from 2 to 26.5 GHz
- Sweep Modes: Full Sweep, F1 to F2 Sweep, M1 to M2 Sweep, Delta F F0 Sweep, Delta F F1 Sweep
- Markers: Video, RF, and Intensity
- Leveling Modes: Internal, Detector, Power Meter
Key Features
- Phase-lock capability to external counter for enhanced frequency accuracy
- External sweep input: 0 to 10 V ramp signal via rear panel BNC connector (10 kΩ impedance)
- External FM and phase lock input via rear panel BNC connector
- Sweep triggering: AUTO, LINE, EXTERNAL (4 to 25 Vpk or TTL-compatible pulse, >1 µsec width, >5 µsec fall time), and SINGLE modes
- Self-test function with diagnostic codes upon power-up or manual activation
- Front panel controls accessible via optional GP-IB interface (Option 03)
Typical Applications Microwave component testing, system alignment, and automated production characterization where frequency stability, signal purity, and precise power control are essential.
Compatibility & Integration
- GP-IB (IEEE-488/IEC 625) Interface: Optional (Option 03) for remote control and automated testing
- Option 2: 70 dB Attenuator for extended power range
- Part of the Wiltron 6600A Series of programmable sweep generators
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz to 18.6 GHz | Power Range (dBm): >10 dBm (internally leveled) |
| >−82 dBm (with Option 2) | Voltage / Current Range | 0 to 10 V (external sweep input) |
Wiltron 6647A-40 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Wiltron 6647A-40. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Wiltron 6647A-40 faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Wiltron 6647A-40 price
This Wiltron 6647A-40 is listed at $4,480. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Wiltron 6647A-40 calibration cost
NIST-traceable calibration for the Wiltron 6647A-40 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →