Anrtisu 68069B Synthesized CW Generator, 0.01 – 40.0 GHz

Anrtisu 68069B Synthesized CW Generator, 0.01 – 40.0 GHz,Provides low-noise, synthesized continuous wave RF signals from 10 MHz to 40 GHz. Used in research and development.

Model
Anritsu 68069B
Calibration
Available on request
Price
$33,580
Status
In stock
RFQ ref
68069B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Anritsu 68069B is a synthesized continuous-wave (CW) signal source covering 0.01 GHz to 40.0 GHz with microprocessor-based frequency control and high-resolution phase-lock capability. It delivers output power from –125 dBm to +13 dBm (extending to +17 dBm across select frequency bands) with 0.01 dB resolution and 2 dB output accuracy. Ultra-low SSB phase noise of –100 dBc at 10 kHz offset from 10 GHz enables precision local oscillator and test applications.


The instrument generates discrete CW frequencies with 1 kHz minimum step size, optionally 0.1 Hz with Option 11, and supports broad and narrow band sweep modes with up to 10,000 steps, variable dwell times from 1 ms to 99 seconds, and step switching under 15 ms + 1 ms/GHz or 40 ms maximum. Spurious signals remain typically –40 dBc for harmonics and non-harmonics. Internal modulation includes AM, FM, phase modulation, and pulse modulation for radar testing; user-downloadable complex modulation and frequency agility up to 1000 non-sequential frequencies under GPIB control are supported.


All functions are accessible via front panel or IEEE-488 GPIB remote control. Harmonic output and alternate sweep modes provide flexible signal generation for R&D, ATE, and frequency conversion applications.


Technical Specifications



  • Frequency Range: 0.01 GHz to 40.0 GHz

  • Output Power: –125 dBm to +13 dBm (up to +17 dBm in select bands); 0.01 dB resolution; ±2 dB accuracy

  • Phase Noise (SSB): –100 dBc at 10 kHz offset from 10 GHz carrier; –96 dBc/Hz to –60 dBc/Hz range

  • Frequency Resolution: 1 kHz standard; 0.1 Hz with Option 11

  • Spurious Signals: Harmonics and non-harmonics typically –40 dBc

  • Step Sweep: 1 to 10,000 steps; dwell time 1 ms to 99 seconds; switching <15 ms + 1 ms/GHz or 40 ms max

  • Frequency Agility: 1000 non-sequential frequencies storable and addressable via GPIB

  • Output Impedance: 50 Ω


Key Features



  • Internal AM, FM, phase, and pulse modulation with advanced radar pulse capability

  • User-defined and downloadable complex modulation

  • Alternate sweep mode with independent frequency and power ranges per sweep

  • Manual stepped frequency sweep with phase-lock adjustment

  • IEEE-488 GPIB full remote control (excluding power on/standby)


Typical Applications



  • Local oscillator replacement

  • Automated test equipment (ATE) and system integration

  • Research and development RF signal sourcing

  • Radar and microwave component testing


Compatibility & Integration IEEE-488 GPIB interface enables remote programmability and integration into automated measurement systems.

Specifications

ParameterValue
Frequency Range0.01 GHz to 40.0 GHz | Power Range (dBm): –125 dBm to +13 dBm (up to +17 dBm in select bands)

Anritsu 68069B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Anritsu 68069B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Anritsu 68069B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Anritsu 68069B price

This Anritsu 68069B is listed at $33,580. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Anritsu 68069B calibration cost

NIST-traceable calibration for the Anritsu 68069B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents