Agilent/Keysight 8146A – Optical Time Domain Reflectometer
The Agilent/Keysight 8146A is a versatile optical measurement instrument designed for characterizing optical components and systems. It offers high accuracy and reliability in measuring optical power, wavelength, and polarization. Ideal for R&D and manufacturing environments.
- Model
- Agilent / Keysight 8146A
- Calibration
- Available on request
- Price
- $1,399
- Status
- In stock
- RFQ ref
- 8146A
This unit
The Agilent/Keysight 8146A is a optical time domain reflectometer (OTDR) engineered for installation, field maintenance, and benchtop characterization of optical fiber systems and components. The instrument automatically analyzes optical links, detecting up to 100 reflective and non-reflective events in under seven seconds with measurement repeatability and precision suitable for R&D, manufacturing, and field service applications.
Technical Specifications
Distance and Resolution- Span: 1 to 200 km selectable
- Read-out resolution: 0.1 m / 0.0001 mi / 0.1 ft
- Distance accuracy: ±0.5 m ± sample spacing
- Sampling resolution: < 5 cm
- Event dead zone: as low as 3 m (module dependent)
- Attenuation dead zone: 30 m or 40 m (module dependent) Dynamic Range & Sensitivity
- Dynamic range: up to 30 dB (module dependent)
- Minimum detectable splice loss: 0.05 dB Physical & Power
- Dimensions: 190 mm H × 340 mm W × 465 mm D
- Net weight: 15 kg including one module
- Power requirement: 120 V
- Laser safety: Class 1 (IEC 60825-1)
Key Features
- Plug-in module architecture supports single-mode and multimode fiber testing
- Expert and Easy operating modes accommodate varying user proficiency levels
- Simultaneous dual-waveform display for comparative analysis
- Comprehensive event detection: splice loss, connector loss, reflection points, and fiber attenuation quantification
- High-speed characterization: complete link analysis in under 7 seconds
Typical Applications
- Installation verification and acceptance testing of optical fiber networks
- Field troubleshooting and maintenance of installed links
- Manufacturing quality assurance and fiber characterization
- Splice and connector loss validation
- Optical return loss and link length estimation
Compatibility & Integration Single-Mode Modules
- HP 81462SH: 1310 nm ±20 nm
- HP 81463SH: 1550 nm ±20 nm
- HP 81465SH: 1310/1550 nm ±20 nm Multimode Modules
- HP 81461MH: 850 nm ±10 nm
- HP 81462MH: 1300 nm ±10 nm
- HP 81464MH: 850/1300 nm ±10 nm
Agilent / Keysight 8146A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 8146A. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Agilent / Keysight 8146A faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Agilent / Keysight 8146A price
This Agilent / Keysight 8146A is listed at $1,399. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight 8146A calibration cost
NIST-traceable calibration for the Agilent / Keysight 8146A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →