HP / Agilent 81536A POWER SENSOR MODULE

The HP/Agilent 81536A is a power sensor module designed for optical power measurements. It offers high accuracy and sensitivity, making it suitable for a variety of applications in optical communication and testing. This module is compatible with HP/Agilent optical power meters and is ideal for characterizing optical components and systems.

Model
HP / Agilent 81536A
Calibration
Available on request
Price
$547
Status
In stock
RFQ ref
81536A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 81536A Power Sensor Module is a plug-in optical power measurement instrument for HP/Agilent lightwave multimeter mainframes. It delivers precise power readings across the near-infrared spectrum using an InGaAs sensor element, making it essential for telecommunications and photonics characterization work. The module inserts into compatible 8153A, 8163A, and 8166A mainframes and provides calibrated measurements with ±2.5% uncertainty across the 1000–1650 nm window.

Technical Specifications

Sensor Element: InGaAs (Indium Gallium Arsenide)

Wavelength Range: 800 nm to 1700 nm

Power Measurement Range: +3 dBm to −70 dBm

Uncertainty (Reference Conditions): ±2.5% over 1000–1650 nm; ±5% ±50 pW total

Linearity: ±0.015 dB ±30 pW (19–38 °C, 0 to −50 dBm); ±0.05 dB ±50 pW (0–55 °C)

Noise (1200–1600 nm, 1 sec average): <50 pW peak-to-peak

Display Resolution: 0.001 dBm or 0.0001 dBm (printout); 100 pW watt resolution

Applicable Fiber: 9/125 µm to 100/140 µm, NA ≤ 0.3

Analog Output: DC to 700 Hz or DC to 4000 Hz depending on range; 0–2 V into open circuit; 600 Ω typical impedance

Warm-up Time: 20 minutes

Recalibration Interval: 1 year

Key Features

InGaAs photodetector optimized for telecom and datacom wavelengths

Wide dynamic range measurement: +3 dBm to −70 dBm

Analog output for real-time signal monitoring and external logging

HP-IB interface support (SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2)

Operating temperature range 0–55 °C; storage to −40–+70 °C

Compact module footprint: 89 mm (H) × 212.3 mm (W) × 355 mm (D)

Typical Applications

Fiber-optic power monitoring in production test

Optical component and subsystem characterization

Wavelength-dependent power measurements across 800–1700 nm

Return loss and insertion loss validation

Compatibility & Integration Inserts directly into HP/Agilent 8153A, 8163A, and 8166A lightwave multimeter mainframes. Operates with standard Agilent optical connectors (FC/PC, SC/PC). Optional 12–30 V DC input available (Option 050).

HP / Agilent 81536A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 81536A. Calibration covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 81536A faults: FC/PC connector damage, EEPROM failure (mainframe shows unrecognized sensor), reduced responsivity. Common faults include FC/PC input connector damage, EEPROM failure causing the module to appear unrecognized by the mainframe, and reduced detector responsivity from contamination.

HP / Agilent 81536A price

This HP / Agilent 81536A is listed at $547. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 81536A calibration cost

NIST-traceable calibration for the HP / Agilent 81536A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. Turnaround is 1 to 3 business days after receipt. See full calibration details →