Agilent Keysight 8156A-100 High Power Optical Attenuator

Agilent / Keysight Optical Attenuators. Resolution: 0.001 dB.

Model
Agilent / Keysight 8156A-100
Calibration
Available on request
Price
$2,789
Status
In stock
RFQ ref
8156A-100
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent/Keysight 8156A-100 is a optical attenuator for precise control of optical power in telecommunications testing and measurement. Operating across 1200 nm to 1650 nm, it delivers 60 dB attenuation range with 0.001 dB resolution and ±0.1 dB accuracy. The instrument accommodates single-mode and 50 µm multimode fiber via external connector interfaces, supporting both R&D and production environments.


A novel single filter design eliminates dark spots and attenuation overshoots or undershoots, ensuring stable performance without ranging artifacts. Repeatability reaches ±0.005 dB typical, while polarization sensitivity remains below 0.075 dBp-p. Insertion loss measures 4.5 dB typical with return loss exceeding 35 dB.


Maximum input power is +23 dBm, with switching time ranging from 20 ms to 400 ms.


Technical Specifications



  • Wavelength Range: 1200 nm to 1650 nm

  • Attenuation Range: 60 dB

  • Resolution: 0.001 dB

  • Attenuation Accuracy (Option 100): ±0.1 dB typical

  • Attenuation Accuracy (Options 101/201): ±0.05 dB typical

  • Polarization Sensitivity (Option 100): <0.075 dBp-p typical

  • Polarization Sensitivity (Options 101/201): <0.02 dBp-p typical

  • Repeatability: 35 dB (Option 100), >45 dB (Options 101/121), >60 dB (Options 201/221), >22 dB (Option 350)

  • Maximum Input Power: +23 dBm

  • Switching Time: 20 ms to 400 ms

  • Operating Temperature: 0°C to 55°C

  • Storage Temperature: -40°C to 80°C

  • Humidity: <95% RH (0°C to 40°C)

  • Power Requirements: 100/120/220/240 Vrms (±10%), 45–400 Hz, 90 VA max

  • Dimensions: 9 mm H × 212.3 mm W × 345 mm D

  • Weight: 5.3 kg (net)


Key Features



  • Single filter design - no dark spots, overshoots, or undershoots

  • Fiber support for single-mode and 50 µm multimode (Option 350)

  • HP-IB/GPIB remote control interface

  • Programmable back reflector mode with external reference reflector (HP 81000BR) compatibility

  • Optional 13 dB monitor tap output on select options (121, 221)

  • Multiple configuration options tailoring return loss and insertion loss performance


Typical Applications



  • Fiber-optic component characterization and testing

  • Telecommunications system validation and debug

  • Optical power level regulation in test benches

  • Reflection sensitivity measurement and analysis

  • Production and manufacturing test environments


Compatibility & Integration The 8156A operates as a mainframe and requires external optical connector interfaces from the Agilent 81000FI series (FC/PC, SC/APC, ST, and other types available). Integration with HP 81000BR reflectors enables back reflector measurements. Multiple performance options - 100, 101, 121, 201, 221, and 350 - allow customization of return loss, insertion loss, monitoring capability, and fiber type support.

Agilent / Keysight 8156A-100 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 8156A-100. Calibration covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / Keysight 8156A-100 faults: motor and gear mechanism wear (attenuation steps or loss of smooth control), optical port contamination, EEPROM calibration data corruption. Most common faults are motor or gear mechanism wear on variable attenuators causing attenuation steps or loss of smooth control, optical port contamination reducing dynamic range, and EEPROM calibration data corruption on digitally controlled models.

Agilent / Keysight 8156A-100 price

This Agilent / Keysight 8156A-100 is listed at $2,789. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight 8156A-100 calibration cost

NIST-traceable calibration for the Agilent / Keysight 8156A-100 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →