HP / Agilent 83624A Synthesized Sweeper, 2 to 20 GHz, High Power,+20 dBm
The HP/Agilent 83624A is a synthesized sweeper that delivers high performance and reliability for demanding applications. Covering a frequency range of 2 to 20 GHz with a high power output of +20 dBm, it provides precise and versatile signal generation for testing and measurement needs.
- Model
- HP / Agilent 83624A
- Calibration
- Available on request
- Price
- $6,444
- Status
- In stock
- RFQ ref
- 83624A
This unit
The HP/Agilent 83624A is a synthesized sweeper delivering signal generation across 2 to 20 GHz with +20 dBm maximum leveled output power. Built for RF and microwave test environments, this instrument combines broadband frequency coverage with indirect synthesis architecture to achieve low single-sideband phase noise of -76 dBc at 10 kHz offset (20 GHz). Standard 1 kHz frequency resolution steps down to 1 Hz with Option 008, supporting three discrete frequency bands with switching times from 5 ms (within-band steps ≤100 MHz) to 70 ms maximum across band boundaries.
Technical Specifications
Frequency & Timing- Frequency Range: 2 to 20 GHz
- Frequency Resolution: 1 kHz (standard); 1 Hz (Option 008)
- Band 1: 2.0 to <7 GHz | Band 2: 7 to 80 dB on/off ratio; <10 ns rise/fall times (Option 006); 300 ns to 400 ms period; 25 ns resolution
- Amplitude Modulation: 100%/V linear or 10 dB/V scan; up to 99.7% depth (50 dB)
- Frequency Modulation: ±8 MHz max deviation; rates to 8 MHz Environmental
- Operating Temperature: 0 to 55°C
Key Features
- Precise 1 Hz frequency resolution enables fine-grained tuning in demanding measurement scenarios
- Fast switching within frequency bands (5 ms for steps ≤100 MHz) accelerates test sequences
- Decoupled AM and FM capabilities allow simultaneous, independent modulation
- Phase noise floor of -76 dBc at 10 kHz supports sensitive receiver characterization
Typical Applications
- Component and subsystem characterization across X, Ku, and K bands
- Transmitter and receiver testing with configurable power and modulation
- System integration and benchtop RF measurement
Compatibility & Integration The 83624A integrates standard swept architecture with option-based enhancements (Option 001 for leveled output; Option 006 for fast pulse edges; Option 008 for 1 Hz resolution). Internal time base accuracy and stability meet precision test requirements without external references.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 2 to 20 GHz | Power Range (dBm): '-20 to +20 dBm |
HP / Agilent 83624A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83624A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 83624A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
HP / Agilent 83624A price
This HP / Agilent 83624A is listed at $6,443.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 83624A calibration cost
NIST-traceable calibration for the HP / Agilent 83624A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →