Fluke 845AB High Impedance Voltmeter – Null Detector

The Fluke 845AB is a high impedance voltmeter and null detector, ideal for precision voltage measurements and sensitive circuit testing. It combines accurate voltage readings with null detection capabilities.

Model
Fluke 845AB
Calibration
Available on request
Price
$1,100
Status
In stock
RFQ ref
845AB
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Fluke 845AB is a precision high-impedance DC voltmeter and null detector engineered for sensitive voltage measurements in laboratory, research, and critical circuit testing environments. It combines a wide measurement range of 1 µV to 1000 V DC across 19 ranges with exceptional input impedance to minimize circuit loading and deliver accurate readings. The instrument functions as both a precision voltmeter and null detector with 0.1 µV/V sensitivity, making it ideal for applications requiring detection of small voltage changes or null conditions in bridge circuits and similar setups.


Technical Specifications



  • Measurement Range: 1 µV to 1000 V DC, 19 ranges

  • Null Detector Sensitivity: 0.1 µV/V

  • Accuracy: ±(2% of end scale + 1 µV)

  • Input Resistance: 100 MΩ (300 mV range and above); 10 MΩ (100 mV range and below)

  • Input Isolation: Better than 10¹² Ω

  • DC Common Mode Rejection: Better than 160 dB

  • AC Normal Mode Rejection (60 Hz and above): 60 dB above end scale produces less than 2% of end scale error

  • AC Common Mode Rejection (below 100 kHz): 120 dB or 100 V rms, whichever is less, produces less than 2% of end scale error

  • Maximum Meter Noise: 0.20 µV (1 µV and other ranges); 0.25 µV (3 µV range)

  • Response Time to 90% Reading: 1.5 to 5 seconds depending on range

  • Zero Stability: Better than 0.15 µV/hour; better than 0.3 µV/day

  • Temperature Coefficient of Zero: Less than 0.1 µV/°C (15°C to 35°C); less than 0.2 µV/°C (0°C to 50°C)

  • Overload Protection: Up to 1100 V DC on any range


Key Features



  • Recorder output: 0–1 V DC, 5 kΩ to 7.5 kΩ source impedance, 0.5% linearity

  • Configurable as DC amplifier with maximum gain of 120 dB

  • Zero control range: ±5 µV minimum

  • Dual-power operation: 115/230 VAC ±10%, 50 to 440 Hz, or rechargeable battery

  • Battery runtime: Approximately 40 hours per full charge

  • Full recharge time: 16 hours

  • Power consumption during recharge: Approximately 6 watts

  • Trickle-charge capability during line operation


Typical Applications



  • Bridge balance and null detection

  • Thermoelectric measurements

  • Potentiometric circuits

  • High-impedance sensor readout

  • Standards calibration and verification

  • Research applications requiring sub-microvolt sensitivity

Specifications

ParameterValue
Voltage / Current Range1 µV to 1000 V DC

Fluke 845AB — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Fluke 845AB. Calibration covers calibration scope varies by instrument type: vacuum gauges are verified at multiple pressure points against NIST-traceable pressure standards; spectrophotometers at wavelength accuracy and photometric accuracy points; strain gauge indicators at bridge excitation accuracy and full-scale output accuracy; pH meters against certified buffer solutions at pH 4, 7, and 10.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Fluke 845AB faults: vacuum gauge filament failure (ionization gauges), spectrophotometer D2 lamp end-of-life, wavelength motor failure, strain indicator front-panel binding post corrosion, pH electrode conditioning failure. Most specialty laboratory instruments fail at their electromechanical interfaces first — vacuum gauge connectors and filament assemblies, spectrophotometer lamp housings and motor drives, strain indicator binding post corrosion. Component-level repair is standard; manufacturer support for legacy instruments is often limited, making third-party bench repair the practical path.

Fluke 845AB price

This Fluke 845AB is listed at $1,100. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Fluke 845AB calibration cost

NIST-traceable calibration for the Fluke 845AB is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers calibration scope varies by instrument type: vacuum gauges are verified at multiple pressure points against NIST-traceable pressure standards; spectrophotometers at wavelength accuracy and photometric accuracy points; strain gauge indicators at bridge excitation accuracy and full-scale output accuracy; pH meters against certified buffer solutions at pH 4, 7, and 10.. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents