Agilent/Keysight 8554B
The Agilent/Keysight 8554B is a spectrum analyzer plug-in designed for use with compatible Agilent/Keysight mainframes. It extends the capabilities of the mainframe to provide spectrum analysis functionality. Key features include frequency range, high sensitivity, and precise measurement capabilities, making it suitable for various RF and microwave applications.
- Model
- HP 8554B
- Calibration
- Available on request
- Price
- $2,699
- RFQ ref
- 8554B
This unit
The Agilent/Keysight 8554B is a spectrum analyzer plug-in module that extends the frequency analysis capabilities of compatible mainframe units. It delivers precise signal component measurements across 100 kHz to 1250 MHz, making it essential for RF and microwave characterization where detailed spectral data and measurement accuracy are critical.
Technical Specifications
Frequency Coverage & Input- Operating range: 100 kHz to 1250 MHz
- Input impedance: 50 Ω nominal
- Maximum input level: +13 dBm peak/average power (1.4 Vac peak) or ±50 Vdc incident on mixer; +33 dBm incident on input attenuator at 50 dB position Amplitude Measurement
- Absolute amplitude calibration range: -122 dBm to +10 dBm (log)
- Amplitude accuracy: ±1.75 dB with proper technique
- Average noise floor: <-102 dBm with 10 kHz IF bandwidth
- Dynamic range supports residual responses 65 dB below -40 dBm input signal IF Bandwidth & Selectivity
- IF bandwidths: 100 Hz to 200 kHz selectable
- 60 dB/3 dB ratio: <20:1 (10 kHz–200 kHz); <11:1 (100 Hz–3 kHz) Stability & Noise Performance
- Residual FM (stabilized): <100 Hz peak-to-peak
- Residual FM (unstabilized): 70 dB below CW signal at 50 kHz or greater offset (1 kHz IF bandwidth) Calibration & Reference Level
- System frequency response: ±1.50 dB
- Log reference level: 70 dB range (60 dB below 200 kHz) in 10 dB steps, ±0.2 dB accuracy
- Vernier control: continuous 12 dB range, ±0.1 dB at 0, -6, -12 dB positions; ±0.25 dB elsewhere
- Calibrator output: -30 dBm (±0.3 dB) at 30 MHz (±3 kHz)
- Input attenuator: 0 to 50 dB in 10 dB steps Display
- Persistence: 0.2 to 60+ seconds variable
- Stored display retention: >2 hours at reduced intensity; 1 minute at normal intensity
- Erase time: 350 milliseconds Physical & Environmental
- Dimensions: 102 mm H × 226 mm W × 344 mm D (4″ × 8.9″ × 13.5″)
- Weight: 17 lb
- Operating temperature: 0° to +55°C
- Storage temperature: -40° to +75°C
Key Features
- Plug-in architecture for Agilent/Keysight mainframe integration
- High-resolution spectral analysis with <-102 dBm noise floor capability
- Precision amplitude calibration across 132 dB dynamic range
- Compact form factor with extended display persistence for signal capture and retention
Typical Applications Frequency response characterization, RF signal detection, spectrum monitoring, and microwave component analysis across the VHF and UHF bands.
Compatibility & Integration Designed as a plug-in module for compatible Agilent/Keysight mainframe units, including the 141T Spectrum Analyzer Mainframe.
HP 8554B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP 8554B. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
HP 8554B price
This HP 8554B is listed at $2,699. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP 8554B calibration cost
NIST-traceable calibration for the HP 8554B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →