Agilent / Keysight E1437A VXI 20 MSa/s 23-bit Digitizer with Filters and Memory

The Agilent/Keysight E1437A is a VXI 20 MSa/s 23-bit Digitizer equipped with filters and memory, designed for high-precision data acquisition. It offers exceptional resolution and sampling rates, making it suitable for demanding applications requiring accurate signal analysis. The built-in filters enhance signal integrity by reducing noise and unwanted frequencies, while the onboard memory facilitates efficient data storage and retrieval.

Model
Agilent / Keysight E1437A
Calibration
Available on request
Price
$5,972
Status
In stock
RFQ ref
E1437A
View cart

This unit

The Agilent/Keysight E1437A is a C-size, single-slot VXI digitizer engineered for high-precision signal acquisition and analysis. Operating at 20 MSa/s with 23-bit ADC resolution, it captures complex signal details required for advanced digital signal processing applications. The module delivers spurious-free dynamic range to -110 dBfs, linearity to 110 dBfs, and noise density of -140 dBfs/Hz at full resolution.


Technical Specifications

Digitizer Performance

  • Resolution: 23-bit raw ADC

  • Sample rate: 20 MSa/s

  • SFDR: Up to -110 dBfs (18 bits effective)

  • Noise density: -140 dBfs/Hz

  • Linearity: Up to 110 dBfs at 20 MSa/s

  • Spurious signal contamination: At least -110 dBfs

  • Real-time distortion correction: Fixed nonlinearities reduced to ≥110 dB below full scale; soft distortion to ≥70 dB below applied carrier level Input and Signal Conditioning

  • Anti-alias filter: 8 MHz

  • Input bandwidth: 8 MHz (filtered), 40 MHz (unfiltered)

  • Digital filters: 24 real-time decimation type

  • Digital local oscillator: Included

  • Analog signal conditioning: Built-in Memory and Data Transfer

  • FIFO memory: 8 MB standard (64 MB optional)

  • VXIbus port transfer rate: 1 MB/s typical

  • Local bus transfer: Up to 60 MB/s burst; 40.96 MB/s continuous Timing and Synchronization

  • Sample clocks: Two standard

  • Internal clock: 20 MHz, 20.48 MHz

  • External reference clock: 20 MHz phase-lockable

  • Flexible triggering and synchronization


Key Features



  • Single-slot C-size VXI module for compact system integration

  • 23-bit resolution with distortion correction for enhanced signal fidelity

  • Dual memory options accommodate varying acquisition duration requirements

  • High-speed local bus enables sustained 40+ MB/s data transfer

  • 24 programmable filters enable in-module signal conditioning and bandwidth optimization


Typical Applications



  • Signal monitoring in digital signal processing workflows

  • Multi-channel data acquisition systems

  • RF and baseband signal analysis

  • Test and measurement environments requiring high-resolution transient capture


Compatibility & Integration VXI plug-and-play architecture. Supported on Windows 3.1, Windows 95, Windows NT, Windows 2000, and HP-UX platforms.