Agilent (HP) E6009A Multimode OTDR Module
Agilent (HP) E6009A,"A multimode OTDR module for locating faults and measuring fiber lengths in multimode networks. Provides high resolution and event detection capabilities. Supports field deployment in LAN and short-haul environments. Easy integration with modular platforms for fiber testing and troubleshooting across telecom infrastructure."
- Model
- Agilent / HP E6009A
- Calibration
- Available on request
- Price
- $130
- Status
- In stock
- RFQ ref
- E6009A
This unit
The Agilent (HP) E6009A Multimode OTDR Module characterizes and locates faults in multimode fiber optic networks operating in LAN and short-haul environments. This module integrates with Agilent's modular OTDR platforms to deliver high-resolution trace acquisition, precise event detection, and accurate fiber length and loss measurements across dual wavelengths.
Technical Specifications
Optical Performance- Wavelengths: 850 nm ± 30 nm and 1300 nm ± 30 nm
- Fiber Type: 62.5 µm multimode
- Pulsewidths: 850 nm (5 ns to 100 ns selectable); 1300 nm (5 ns to 1 µs selectable)
- Dynamic Range: 850 nm 12/12 dB (10 ns/100 ns); 1300 nm 18/18 dB (10 ns/100 ns)
- Event Dead Zone: 3 m
- Attenuation Dead Zone: 10 m
- Reflectance Threshold: ≤ -35 dB
- Distance Accuracy: ± (1 m offset + 10⁻⁴ scale factor × distance + 0.5 sampling spacing)
- Signal-to-Noise Ratio: ≥ 15 dB (1 µs pulsewidth, 3 minutes averaging)
- Event Detection Thresholds: Non-reflective (0.0–5.0 dB, 0.01 dB steps); Reflective (-14.0 to -65.0 dB, 0.1 dB steps); Fiber breaks (0.1–10 dB, 0.1 dB steps); Fiber end (0.1–20 dB, 0.1 dB steps) Data Acquisition
- Measurement Points: Up to 16,000 points
- Trace Format: Bellcore/Telcordia OTDR (GR 196 Rev. 1.0/1.1; SR-4731 Rev. 2.0)
- Storage: 3.5" Floppy Disk (720/1440 kByte); PCMCIA Type II Memory Card (440 MB, 13,000 traces typical) Interfaces & Connectivity
- RS232C: 115,200 bps (4,000 points ≈ 1 second; 16,000 points ≈ 4 seconds)
- Centronics: Standard parallel port (SPP)
- Keyboard: PS2 (Min-DIN) compatible
- Fiber Connectors: SC (81000KI) and ST (81000VI) supported Environmental
- Operating Temperature: 0°C to 50°C (5°C to 45°C for floppy disk operation)
Key Features
- Dual-wavelength operation for comprehensive multimode fiber analysis
- Selectable pulse widths optimize resolution versus dynamic range trade-offs
- Fine-grained event detection thresholds enable precise fault identification
- Bellcore/Telcordia trace compliance ensures standards alignment
- Multiple storage options support field and lab workflows
Typical Applications
- Local area network (LAN) fiber characterization
- Short-haul link fault location and distance measurement
- Multimode fiber splice and connector analysis
- Installation acceptance testing
Compatibility & Integration The E6009A integrates into Agilent modular OTDR platforms. RS232C and Centronics interfaces enable remote operation and data transfer. SC and ST fiber connectors accommodate standard deployment scenarios.
Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 850 nm ± 30 nm and 1300 nm ± 30 nm |
Agilent / HP E6009A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / HP E6009A. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Agilent / HP E6009A faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Agilent / HP E6009A price
This Agilent / HP E6009A is listed at $130. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / HP E6009A calibration cost
NIST-traceable calibration for the Agilent / HP E6009A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →