Agilent HP Keysight E8257N PSG Analog Sweep Generator

Agilent HP Keysight E8257N PSG Analog Sweep Generator,High-performance analog signal generator delivering precise and stable RF signals up to 50 GHz.

Model
Agilent / HP / Keysight E8257N
Calibration
Available on request
Price
$34,570
Status
In stock
RFQ ref
E8257N
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent HP Keysight E8257N PSG is a fully synthesized analog signal generator delivering precision RF output across multiple configurable frequency bands. Built for demanding RF and microwave test applications, it combines wide frequency coverage, fine frequency resolution, and flexible modulation to support component characterization, system integration, and production test environments.


Technical Specifications

Frequency Range & Resolution
The E8257N offers six frequency band options: 250 kHz to 20 GHz (Option 520), 10 MHz to 20 GHz (Option 521), 250 kHz to 31.8 GHz (Option 532), 250 kHz to 40 GHz (Option 540), 250 kHz to 50 GHz (Option 550), and 250 kHz to 67 GHz (Option 567). All bands operate down to 100 kHz. Frequency resolution is 0.001 Hz across all modes. Output Power
Standard RF output spans −105 dBm to +10 dBm at 18 GHz and +7 dBm at 40 GHz. High output power and ultrahigh output power options are available. Maximum power is warranted from 15 to 35 °C and typical from 0 to 15 °C, with degradation less than 2 dB at 35 to 55 °C. Spectral Purity
Phase noise measures typically −126 dBc/Hz at 10 GHz carrier with 10 kHz offset. Enhanced phase noise options (UNR, UNX, UNY) and improved harmonics below 2 GHz (Option 1EH) are available.

Key Features



  • Analog modulation: AM, FM, phase modulation, and LF output (Option UNT)

  • Pulse and narrow pulse modulation (Options UNU, UNW)

  • Step and ramp sweep with 1 ms to 60 s sweep time

  • Ten independent continuously variable frequency markers

  • Frequency switching within 0.1 ppm (above 250 MHz) or 100 Hz (below 250 MHz)

  • CW switching speed ≥5 ms nominal; 30–40 ms with Option HY2

  • 45-minute warm-up time required

  • Operating temperature: 0 to 50 °C


Typical Applications Component and subsystem characterization, receiver sensitivity testing, transmitter validation, and production alignment across microwave and millimeter-wave bands.


Compatibility & Integration Front panel S-type connectors for RF output and control interfaces support GPIB triggering and marker functionality.

Specifications

ParameterValue
Frequency Range250 kHz to 67 GHz (Option 567)
operation down to 100 kHz on all options | Power Range (dBm)−105 dBm to +10 dBm at 18 GHz

Agilent / HP / Keysight E8257N — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / HP / Keysight E8257N. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / HP / Keysight E8257N faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Agilent / HP / Keysight E8257N price

This Agilent / HP / Keysight E8257N is listed at $34,570. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / HP / Keysight E8257N calibration cost

NIST-traceable calibration for the Agilent / HP / Keysight E8257N is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents