Instek GOM-802 DC Milli-Ohm Meter

The Instek GOM-802 is a high-precision DC milli-ohm meter designed for accurate low-resistance measurements. It features a wide measurement range, high resolution, and excellent stability. Ideal for testing contact resistance, coil resistance, and other low-resistance components.

Model
Instek GOM-802
Calibration
Available on request
Price
$450
Status
In stock
RFQ ref
GOM-802
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Instek GOM-802 is a precision DC milli-ohm meter designed for low-resistance measurement across a 30 mΩ to 3 MΩ range with 0.05% accuracy and 1 µΩ resolution. Built on four-terminal Kelvin measurement technique, it delivers stable results across demanding test environments including contact resistance verification, coil evaluation, and material conductivity assessment.

Technical Specifications

Display: 30,000 counts

Accuracy: 0.05%

Measurement Range: 30 mΩ to 3 MΩ

Resolution: 1 µΩ

Test Current: 1 µA to 1 A (range-dependent: 10 mA at 30 mΩ, 100 mA at 300 mΩ–30 Ω, 1 mA at 300 Ω, 100 µA at 3 kΩ–30 kΩ, 10 µA at 300 kΩ, 1 µA at 3 MΩ)

Sampling Rate: 7 samples/sec (30,000 counts) or 30 samples/sec (3,000 counts)

Operating Temperature: 18°C to 28°C

Warm-up Time: 30 minutes

Power: AC 100/120/220/230 V, 50/60 Hz

Dimensions: 251(W) × 91(H) × 291(D) mm

Weight: Approximately 3 kg

Key Features

Four-terminal Kelvin measurement eliminates lead resistance errors

Continuous or triggered measurement modes for flexible test workflows

Relative (REL) measurement, actual value, and percentage value measurement functions

Hi/Lo comparator with limit percent setting and Go/NoGo testing capability

PASS/FAIL alarm output with memory storage for 20 sets of Hi/Lo/GO settings

Diode test and temperature measurement with compensation

Manual or autoranging

Auto-recall of last setting after power cycle

Handler interface with PASS, FAIL, HI, LO, READY, and EOT output signals plus trigger input

Typical Applications

Contact resistance testing in switches and relays

Coil resistance evaluation in motors and transformers

Material conductivity assessment

Quality assurance and component validation

Go/NoGo production testing with integrated comparator

Compatibility & Integration Standard: Handler interface with scan support and output/input signal control Optional Interfaces: RS-232C (DB-9 male connector) and GPIB Accessories: PT-100 platinum temperature probe, GTL-232 RS-232C cable, GTL-108A test leads (included)

Specifications

ParameterValue
Voltage / Current Range50/60 Hz

Instek GOM-802 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Instek GOM-802. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Instek GOM-802 faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.

Instek GOM-802 price

This Instek GOM-802 is listed at $450. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Instek GOM-802 calibration cost

NIST-traceable calibration for the Instek GOM-802 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents