HP / Agilent 4349B 4 Channel High Resistance Meter, DC

HP / Agilent Laboratory Measurement Instruments. Accuracy: ±2%.

Model
HP / Agilent 4349B
Calibration
Available on request
Price
$1,260
Status
In stock
RFQ ref
4349B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 4349B is a 4-channel high resistance meter engineered for simultaneous DC resistance measurements across multiple devices under test. Built for high-volume production environments, it delivers resistance characterization from 10 kΩ to 1.0 × 10¹⁵ Ω with 2% basic accuracy and selectable display resolution of 3, 4, or 5 digits. Measurement cycles complete in 9.5 ms, 28 ms, 98 ms, or 397 ms, enabling rapid throughput in automated test sequences.


Technical Specifications



  • Resistance Range: 10 kΩ to 1.0 × 10¹⁵ Ω

  • Current Range: 1 pA to 10 µA

  • Basic Accuracy: ±2%

  • Measurement Times: 9.5 ms, 28 ms, 98 ms, or 397 ms

  • Input Terminal Voltage Rating: ±50 V dc

  • Maximum Test Voltage (external source): 1000 V dc

  • Input/Output Resistance: 1 kΩ

  • Display Resolution: 3, 4, or 5 digits (selectable)

  • Power Input: 90–132 V ac or 198–264 V ac; 47–66 Hz

  • Power Consumption: 45 VA typical

  • Operating Temperature: 0 °C to 45 °C

  • Operating Humidity: ≤95% RH at 40 °C

  • Dimensions: 320 mm (W) × 100 mm (H) × 450 mm (D)

  • Weight: 6.5 kg typical


Key Features



  • Simultaneous 4-channel measurement capability for parallel device testing

  • Auto and manual ranging for resistance measurements

  • Contact check verification (2 ms) for fixture and lead continuity

  • Open error correction for leakage current and capacitance compensation

  • Comparator function with High/In/Low pass/fail criteria

  • Programmable test sequence execution including charge-measure-discharge cycles

  • Save/recall of up to 10 measurement setups

  • External DC power source required; recommended supply specifications include ≤1 mV rms ripple (50/60 Hz), ≤5 µV rms/Hz wideband noise (50 Hz), and ≤50 mV rms switching noise (100 kHz)


Typical Applications Quality assurance in capacitor manufacturing, component characterization in R&D, and leakage current assessment in electronics production. High-speed parallel testing accelerates throughput in high-volume operations.


Compatibility & Integration HP-IB (IEEE 488) interface for remote instrument control and data transfer. Handler interface supports negative logic with isolated signals for High/In/Low, No Contact, EOM, Index, Alarm, Keylock, and external trigger.

Specifications

ParameterValue
Voltage / Current Range±50 V dc (input terminals)

HP / Agilent 4349B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 4349B. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 4349B faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.

HP / Agilent 4349B price

This HP / Agilent 4349B is listed at $1,259.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 4349B calibration cost

NIST-traceable calibration for the HP / Agilent 4349B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →