JDS Uniphase JDSU HA9 HA09+20AFA1 Programmable Attenuator
JDS Uniphase/JDSU Optical Attenuators. Resolution: 0.01 dB, Accuracy: ±0.1 dB (or greater of ±0.1 dB or ±0.004 dB/dB.
- Model
- JDS Uniphase/JDSU HA9/HA09+20AFA1
- Calibration
- Available on request
- Price
- $2,930
- Status
- In stock
- RFQ ref
- HA9/HA09+20AFA1
This unit
The JDS Uniphase JDSU HA9 HA09+20AFA1 is a programmable optical attenuator that delivers precise, dynamic control of optical signal levels across extended attenuation ranges. Engineered for optical communication system testing, power meter calibration, and laboratory applications, this instrument combines broad dynamic range with fine resolution and excellent repeatability to support demanding measurement and validation workflows.
Technical Specifications
- Attenuation Range: Up to 100 dB (60 dB for optional HA9W extended wavelength)
- Attenuation Resolution: 0.01 dB
- Attenuation Repeatability: ±0.01 dB
- Attenuation Accuracy: ±0.1 dB (or greater of ±0.1 dB or ±0.004 dB/dB to ±0.015 dB/dB depending on calibration optimization and wavelength range)
- Attenuation Change Rate: <2.5 seconds for 0–100 dB transition
- Operating Wavelength Range: 1200–1700 nm standard; 750–1700 nm optional (HA9W)
- Insertion Loss: <1.5 dB single-mode; 45 dB single-mode
- Maximum Optical Input Power: 200 mW
- Polarization Dependent Loss (PDL): 0.03 dB typical; 0.08 dB maximum
- Wavelength Dependence: ±0.05 dB over specified ranges
- Beam Block Attenuation: >90 dB
- Power Requirements: 90–240 V AC, 50–60 Hz
- Power Consumption: 80 VA maximum
Key Features
- Built-in beam block switch for rapid access to infinite attenuation
- Calibration and offset functions enable display matching to external optical power meters
- Discrete optical reflections minimized to >60 dB; cavity effects virtually eliminated
- Front panel keyboard control with remote capability via GPIB (IEEE 488.2) and RS-232
- SCPI/HP8156A command compatibility with supplied LabVIEW drivers
- Configurable couplers (2/98, 10/90, 30/70, 50/50) and optional connector adapters
- Rear panel 5 V output for external switch driver integration
Typical Applications
- Optical power meter calibration and verification
- Dynamic range testing in optical communication systems
- Laboratory attenuation measurement and control
- CATV AM system characterization
Compatibility & Integration Software compatibility maintained with HP 8157A attenuator. Remote control via industry-standard GPIB and RS-232 interfaces supports integration into automated test systems.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 50–60 Hz |
| 80 VA maximum, 90–240 V AC | Wavelength Range | 1200–1700 nm (standard) |
JDS Uniphase/JDSU HA9/HA09+20AFA1 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the JDS Uniphase/JDSU HA9/HA09+20AFA1. Calibration covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common JDS Uniphase/JDSU HA9/HA09+20AFA1 faults: motor and gear mechanism wear (attenuation steps or loss of smooth control), optical port contamination, EEPROM calibration data corruption. Most common faults are motor or gear mechanism wear on variable attenuators causing attenuation steps or loss of smooth control, optical port contamination reducing dynamic range, and EEPROM calibration data corruption on digitally controlled models.
JDS Uniphase/JDSU HA9/HA09+20AFA1 price
This JDS Uniphase/JDSU HA9/HA09+20AFA1 is listed at $2,930. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
JDS Uniphase/JDSU HA9/HA09+20AFA1 calibration cost
NIST-traceable calibration for the JDS Uniphase/JDSU HA9/HA09+20AFA1 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →