The Anritsu MG362X1A is a RF/microwave signal generator engineered for R&D, manufacturing, and depot repair environments requiring exceptional signal purity and frequency stability. Operating from 9 kHz to 70 GHz across multiple models, this generator delivers phase noise of −136 dBc/Hz typical at 10 GHz (10 kHz offset), with harmonics better than −58 dBc and spurious levels below −63 dBc. The instrument supports AM, FM, PM, and pulse modulation with simultaneous modulation modes for complex waveforms, including pulse rise/fall times of 5 ns typical and quad pulse signal generation. Frequency resolution extends to 0.001 Hz, with output power ranging from −130 dBm to +20 dBm at 20 GHz. Internal frequency stability options include OCXO (aging < ±5 × 10⁻⁷ per year) and Rubidium (aging < ±1 × 10⁻⁹ per year, Allan deviation < 8.0 × 10⁻¹² over 100 sec). External reference synchronization via GNSS/GPS is available, with reference frequency I/O at 10 MHz, 100 MHz, and 1600 MHz. A 7-inch touch screen interface with four front-panel USB 3.0 ports enables intuitive local and remote operation.

Technical Specifications

Frequency Range: 9 kHz to 70 GHz (operable to 72 GHz); models available at 20 GHz, 43.5 GHz, and 70 GHz

Frequency Resolution: 0.001 Hz

Output Power: −130 to +20 dBm (20 GHz), −130 to +16 dBm (43.5 GHz), −100 to +6 dBm (70 GHz); leveled output to +30 dBm

Harmonics: Better than −58 dBc typical; −55 dBc at 20/43.5/70 GHz

Spurious (Non-Harmonics): Less than −63 dBc

SSB Phase Noise: −136 dBc/Hz typical; −140 dBc/Hz measured at 10 GHz, 10 kHz offset

Modulation Types: AM, FM, PM, Pulse; simultaneous AM/FM, AM/PM, FM/Pulse

Pulse Characteristics: 5 ns rise/fall time typical; leveled pulse modulation, quad pulse, pulse sweep

LF Waveforms: Sine, Square, Pulse, Triangle, Ramp, Gaussian Noise, Uniform Noise, User Defined

Sweep Modes: Analog, Step, List (9 kHz to full range)

Internal Time Base Stability: OCXO (±5 × 10⁻⁷/year aging); Rubidium (±1 × 10⁻⁹/year aging, <8.0 × 10⁻¹² Allan deviation/100 sec)

Reference I/O: 10 MHz, 100 MHz, 1600 MHz

Interface: 7-inch touch screen, keypad, dial; four USB 3.0 front-panel ports

Key Features

Optional high-power output (Option 15)

Low phase noise enhancements (Options 3, 13, 23)

GNSS/GPS atomic clock receiver (Option 66) for external synchronization

Pulse train generator option for radar signal simulation

Typical Applications

Component and subsystem characterization

Radar and electronic warfare testing

Receiver sensitivity and linearity measurements

Phase noise-dependent applications

Complex waveform simulation and validation

Specifications

ParameterValue
Frequency Range9 kHz to 70 GHz | Power Range (dBm): '-130 dBm to +20 dBm (20 GHz)

Anritsu MG362X1A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Anritsu MG362X1A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Anritsu MG362X1A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Anritsu MG362X1A price

This Anritsu MG362X1A is listed at $18,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Anritsu MG362X1A calibration cost

NIST-traceable calibration for the Anritsu MG362X1A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →