Anritsu MU909014C6 – µOTDR Module, 3-Wavelength, 30 dB Class OTDR With PM, PON-PM, LTS And LS For MT9090A Network Master Series Mainframes
Anritsu OTDRs. Dynamic Range: ** Up to 38 dB, Resolution: ** 2 cm.
- Model
- Anritsu MU909014C6
- Calibration
- Available on request
- Price
- $2,800
- Status
- In stock
- RFQ ref
- MU909014C6
This unit
The Anritsu MU909014C6 is a tri-wavelength µOTDR module engineered for the MT9090A Network Master series mainframes. This 30 dB class OTDR delivers integrated optical testing across power measurement, PON diagnostics, light source generation, and loss characterization. The module supports fiber installation and maintenance workflows, from initial deployment through long-term network validation.
Technical Specifications
OTDR Wavelengths: 1310 nm, 1550 nm, and filtered 1625 nm or 1650 nm
Dynamic Range: Up to 38 dB
Resolution: 2 cm
Data Sampling: 2 cm
Dead Zone: Less than 1 m (approximately 0.8 m)
Launch Fiber: Integrated 10 m fiber for connector reference measurements
PON Measurement: 1490 nm (data) and 1550 nm (video) simultaneous dual-wavelength identification in FTTx networks with splits to 1 × 64
Key Features
Fiber Visualizer mode automates parameter selection and delivers graphical fiber summary without OTDR expertise
Integrated Power Meter (PM) and PON Power Meter (PON-PM) with threshold-based PASS/FAIL evaluation
Light Source (LTS) offering continuous-wave and modulated output (270 Hz, 1 kHz, 2 kHz) at 1310 nm and 1550 nm
Loss Test Set (LS) for direct optical loss measurement
User-defined event thresholds for splice loss, reflectance, and total span loss validation
Telcordia SR-4731 data format support ensures compatibility with legacy Anritsu, NetTest, and third-party equipment
Typical Applications FTTx network deployment and maintenance. PON monitoring and signal verification. Fiber event characterization and loss budgeting. Field troubleshooting requiring portable, multi-function optical diagnostics.
Compatibility & Integration
Mainframe: MT9090A Network Master series only
Connectivity: Ethernet, Wi-Fi, Bluetooth (USB dongles may be required)
Form Factor: Handheld, pocket-size design with high-resolution widescreen color display for indoor and outdoor operation
Anritsu MU909014C6 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu MU909014C6. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu MU909014C6 faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Anritsu MU909014C6 price
This Anritsu MU909014C6 is listed at $2,800. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu MU909014C6 calibration cost
NIST-traceable calibration for the Anritsu MU909014C6 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →