Anritsu MU909015C6 – µOTDR Module, 3-Wavelength, 35 dB Class OTDR For MT9090A Network Master Series Mainframes
The Anritsu MU909015C6 is a µOTDR module designed for use with MT9090A Network Master series mainframes. This 3-wavelength OTDR module offers a dynamic range of 35 dB, making it suitable for a variety of optical fiber testing applications.
- Model
- Anritsu MU909015C6
- Calibration
- Available on request
- Price
- $4,100
- Status
- In stock
- RFQ ref
- MU909015C6
This unit
The Anritsu MU909015C6 is a micro optical time-domain reflectometer (µOTDR) module engineered for integration into MT9090A Network Master series mainframes. This single-mode fiber test module delivers three-wavelength OTDR capability with 35 dB dynamic range, enabling precise characterization of optical fiber infrastructure across installation, maintenance, and troubleshooting workflows. The compact design supports premise, access, metro, and core network testing, including FTTx deployments with high split ratios.
Technical Specifications Wavelengths & Accuracy
Triple-wavelength operation: 1310 nm, 1550 nm, and 1625 nm or 1650 nm (variant-dependent)
Alternative variant: 1310 nm, 1490 nm, and 1550 nm
Wavelength accuracy: ±20 nm Dynamic Range & Resolution
35 dB class; up to 38 dB at 20 µs pulse width
Event dead zone: <1 m (typically ≤0.8 m Fresnel)
Backscatter dead zone: <4.5 m (≤4.0 m at 1310 nm; ≤4.5 m at 1490/1550/1625/1650 nm)
Sampling resolution: 2 cm minimum
Sampling points: <250,001 Distance & Accuracy
Range: 0.5 km to 250 km
Distance accuracy: ±1 m ±3 × measurement distance × 10⁻⁵ ± marker resolution
Measurement accuracy: ±0.5 dB
Linearity: ±0.05 dB or ±0.1 dB Operational Parameters
Pulse width: 5 ns to 500 ns; 1 µs to 20 µs
IOR range: 1.3000 to 1.7000
Return loss measurement: 14.8 ± 2 dB
Key Features
Full OTDR modes: automatic, manual, real-time
Integrated power meter, light source, and loss test set functions
PON power meter capability
Fiber visualizer mode for field operations
Visual fault locator support (optional)
Connector inspection with optional video probe
USB 1.1 and Ethernet connectivity; Bluetooth and Wi-Fi
Operating temperature: −10 to 50 °C
Compact form factor: 190 mm × 96 mm × 48 mm, 700 g
Power & Integration
Module powered by MT9090A mainframe battery
AC input: 100 V to 240 V (50/60 Hz); DC: 9 V
NiMH battery standard (~8 hours operation)
Alkaline dry battery option available
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 100-240 VAC (50/60 Hz), 9 VDC | Wavelength Range: 1310 nm, 1490 nm, 1550 nm, 1625 nm or 1650 nm (variant-dependent) |
| alternative | 1310 nm |
Anritsu MU909015C6 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu MU909015C6. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu MU909015C6 faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Anritsu MU909015C6 price
This Anritsu MU909015C6 is listed at $4,100. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu MU909015C6 calibration cost
NIST-traceable calibration for the Anritsu MU909015C6 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →