Ametek / VTI Instruments / VXI Technology SMP3001DS 64-Channel Multiplexer Module
The Ametek/VTI Instruments/VXI Technology SMP3001DS is a 64-channel multiplexer module designed for data acquisition systems. It allows for the sequential sampling of multiple input signals through a single measurement channel, increasing efficiency and reducing system complexity. Ideal for applications requiring the monitoring of numerous sensors or test points.
- Model
- Ametek / VTI Instruments / VXI Technology SMP3001DS
- Calibration
- Available on request
- Price
- $1,200
- Status
- In stock
- RFQ ref
- SMP3001DS
This unit
The SMP3001DS is a 64-channel multiplexer module for VXIbus systems, engineered to route multiple input signals through a single measurement channel. Part of the SMIP II series, this plug-in module delivers high-density switching in a compact form factor, enabling efficient sequential sampling of numerous sensors and test points while reducing system complexity. Configurable as either a 1×64 two-wire multiplexer or 1×128 one-wire multiplexer, the SMP3001DS integrates directly into VXIbus mainframes and combines with other SMIP II modules to build flexible, scalable switch matrices supporting up to 384 two-wire channels per double-slot VXIbus card.
Technical Specifications
- Channels: 64 (1×64 two-wire or 1×128 one-wire)
- Switching Voltage: 300 VDC, 300 VAC
- Switching Current: 2 A maximum per channel
- Switching Power: 60 W DC, 125 VA; 60 W DC, 62.5 VA
- Bandwidth: >50 MHz (−3 dB), >100 MHz
- Path Resistance: <0.5 Ω, 10⁸ Ω
- VXIbus Slots: Single-slot (SMP1100) or double-slot (SMP1200) compatible
Key Features
- Multi-layer PCB construction with ground planes and shielding to minimize crosstalk
- Internal capacitive discharge relays protect sensitive circuits from residual voltages; discharge function is software-disableable for high-throughput, low-voltage applications
- Signal and control grounds isolated to maintain measurement integrity
- Break-Before-Make (BBM) and Make-Before-Break (MBB) switching implemented in hardware
- Programmable scan lists with configurable timing delays and backplane trigger advancement
- Individual relay control and path-level programming options
- Internal confidence checking confirms relay closure status
- Extensive triggering capability for synchronized multi-instrument operation
Typical Applications Multi-point sensor monitoring, automated test systems, data acquisition from distributed measurement nodes, high-density signal switching in laboratory and production environments.
Compatibility & Integration Register-mapped VXIbus interface with VXIplug&play driver support. Mixes and matches with other SMIP II modules within SMP1100 and SMP1200 VXIbus mainframes.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | >100 MHz, >50 MHz | Voltage / Current Range: 2 A maximum per channel, 300 VDC / 300 VAC |