Astronics / EADS / Racal 1260-30B Multiplexer VXI Module
The Astronics/EADS/Racal 1260-30B is a VXI Multiplexer Module designed for signal routing and switching applications. This module allows for the expansion of measurement and control systems by providing multiple channels that can be selectively connected to a common bus. Ideal for automated testing, data acquisition, and signal distribution systems.
- Model
- Astronics / EADS / Racal 1260-30B
- Calibration
- Available on request
- Price
- $1,050
- Status
- In stock
- RFQ ref
- 1260-30B
This unit
The Astronics/EADS/Racal 1260-30B is a C-size VXIbus multiplexer module that provides dual 1×20 two-wire switching for signal routing in automated test and data acquisition systems. This electromechanical relay-based module handles switching up to 250 VDC or 250 VAC rms at currents to 2 A, with bandwidth to 10 MHz. It integrates into standard VXIbus chassis and operates under message-based or register-based control via an Option 01 Smart Card, making it suitable for multi-point scanning, continuity testing, and confidence checking via relay coil current monitoring.
Technical Specifications
Form Factor: C-size, single-slot VXIbus module
Configuration: Dual 1×20 two-wire multiplexer
Switching Voltage: 250 VDC or 250 VAC rms maximum
Switching Current: 2 A maximum
Bandwidth: 10 MHz
Relay Type: Electromechanical
Power Supply: +5 VDC, 150 mA typical
Operating Temperature: 0 °C to 55 °C
Non-operating Temperature: −40 °C to +75 °C
Relative Humidity: 85 % ±5 % non-condensing at 30 °C
Operating Altitude: 10,000 ft maximum
Weight: 3.2 lbs (1.45 kg) without Option 01T
Key Features
Relay coil current monitoring for confidence checking of switching integrity
Mounting holes per channel enable integration of user-supplied components
Register-based and message-based control protocols supported
Compact single-slot design for dense test system configurations
Typical Applications
Multi-point UUT test scanning with DMM or continuity testers
High-voltage and high-current signal routing in automated test systems
RF and audio signal switching within integrated test platforms
Data acquisition point selection across multiple signal types
Compatibility & Integration The 1260-30B requires integration into a VXIbus-compatible chassis and an Option 01 Smart Card configured as system master for control. It complies with EN61010-1 (safety), EN55011A (emissions), and IEC801-2, 3, 4 (immunity) standards. Proper third-wire earth grounding is mandatory for safe operation.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz | Voltage / Current Range: 2 A max |
Astronics / EADS / Racal 1260-30B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Astronics / EADS / Racal 1260-30B. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Astronics / EADS / Racal 1260-30B faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.
Astronics / EADS / Racal 1260-30B price
This Astronics / EADS / Racal 1260-30B is listed at $1,050. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Astronics / EADS / Racal 1260-30B calibration cost
NIST-traceable calibration for the Astronics / EADS / Racal 1260-30B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →