Fluke / Wavetek / Datron 1362S 6.5 Digit Multimeter VXI Module
Fluke/Wavetek/Datron VXI Modules. Resolution: 4.5 to 6.5 digits (selectable), Resolution: 1 µV (5.5 digits), Resolution: 10 µΩ (6.5 digits).
- Model
- Fluke/Wavetek/Datron 1362S
- Calibration
- Available on request
- Price
- $1,650
- Status
- In stock
- RFQ ref
- 1362S
This unit
The Fluke/Wavetek/Datron 1362S is a 6.5-digit VXI multimeter module engineered for precision measurement within automated test systems. This C-size, single-slot VXIbus card delivers selectable resolution from 4.5 to 6.5 digits across DC voltage, AC voltage, resistance, and current measurements. Read rates reach 1000 readings per second at 4.5-digit resolution, stepping down to 5 readings/sec at 6.5-digit resolution for DC voltage. The module stores measurement blocks of 1 to 1000 readings and supports both front-panel and VXI backplane triggering via synchronous and asynchronous protocols. Programming is handled through SCPI-compatible commands over IEEE-488.2 interfaces, enabling seamless integration into complex test architectures.
Technical Specifications Measurement Resolution & Speed
Resolution: 4.5 to 6.5 digits (selectable); up to 6.5 digits for DC voltage and resistance; 5.5 digits for AC current and AC voltage
Read Rate: Up to 1000 readings/sec at 4.5-digit resolution; 5 readings/sec at 6.5-digit resolution (DC voltage)
Data Storage: 1 to 1000 readings per block at high read rates DC Voltage (DCV)
Ranges: 100 mV to 300 V (decade steps)
Resolution: 1 µV (5.5 digits), 100 nV (6.5 digits)
Input Impedance: 10 GΩ (0.1–10 V ranges); 10 MΩ (100–300 V ranges)
CMRR: > 146 dB at DC; > 80 dB + NMRR at 1–60 Hz
NMRR: > 54 dB at 50/60 Hz; adds 20 dB with filter enabled
Overvoltage Protection: 300 V RMS (all ranges)
Accuracy (90 days, 23°C ± 5°C): 100 mV range ± (30 ppmR + 6 ppmFS); 10 V range ± (20 ppmR + 2 ppmFS); 300 V range ± (30 ppmR + 3 ppmFS) True RMS AC Voltage (ACV)
Ranges: 100 mV to 300 V (decade steps)
Resolution: 1 µV (5.5 digits)
Input Impedance: 1 MΩ / 100 pF
CMRR: > 80 dB (DC to 60 Hz)
Crest Factor: 5:1 at full range
Overvoltage Protection: 300 V RMS (all ranges)
Accuracy (90 days, 23°C ± 5°C): 10–40 Hz ± (0.4%R + 0.1%FS); 40 Hz–20 kHz ± (0.035%R + 0.01%FS); 300 kHz–1 MHz ± (2%R + 1%FS) Resistance (Ω)
Ranges: 100 Ω to 10 MΩ (decade steps)
Resolution: 10 µΩ (6.5 digits)
Maximum Lead Resistance: 100 Ω (any or all leads)
Accuracy (90 days, 23°C ± 5°C): 100 Ω range ± (35 ppmR + 6 ppmFS); 10 kΩ range ± (30 ppmR + 3 ppmFS); 10 MΩ range ± (200 ppmR + 4 ppmFS) Integration & Control
VXI Conformance: VXI specification revision 1.4
Programming: SCPI-compatible, IEEE-488.2
Triggering: Front-panel or VXI backplane (synchronous and asynchronous)
Calibration: Programmable external calibration (covers-on)
Typical Applications Automated component testing, precision voltage and resistance validation, high-speed data acquisition in production test environments, and integration into rack-mounted VXI test systems requiring traceable measurement standards.
Compatibility & Integration The 1362S operates as a C-size VXI module compatible with standard VXIbus architectures. SCPI command language ensures cross-platform software integration, while dual triggering modes support synchronized multi-module measurement sequences.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 Hz to 1 MHz | Voltage / Current Range: DC: 100 mV to 300 V |
| AC | 100 mV to 300 V |
| Current (optional) | 1000 mA |
Fluke/Wavetek/Datron 1362S — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Fluke/Wavetek/Datron 1362S. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Fluke/Wavetek/Datron 1362S faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.
Fluke/Wavetek/Datron 1362S price
This Fluke/Wavetek/Datron 1362S is listed at $1,650. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Fluke/Wavetek/Datron 1362S calibration cost
NIST-traceable calibration for the Fluke/Wavetek/Datron 1362S is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →