Keysight / Agilent / HP 4328A Milliohmmeter

The Keysight 4328A Milliohmmeter is a precision instrument designed for accurate low-resistance measurements. It features a wide measurement range, high resolution, and stable performance, making it suitable for various applications including component testing, quality control, and material research. The 4328A offers both manual and automatic ranging, ensuring ease of use and optimal measurement accuracy. Its solid design and reliable operation make it a valuable tool for any electronics lab or production environment requiring precise low-resistance measurements.

Model
Keysight / Agilent / HP 4328A
Calibration
Available on request
Price
$450
Status
In stock
RFQ ref
4328A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Keysight 4328A Milliohmmeter is a precision instrument for low-resistance measurement using Kelvin Bridge methodology and four-terminal measurement technique. It delivers high sensitivity across 0.001 Ω to 100 Ω full scale with maximum resolution of 0.02 Ω, achieving ±2% accuracy. The instrument employs constant applied current with a 1000 Hz ±100 Hz measurement frequency and maintains sample voltage below 200 µV peak at full scale - never exceeding 20 mV peak under any condition, including incorrect range settings.

Technical Specifications

Measurement Range: 0.001 Ω to 100 Ω full scale in 1, 3, 10 sequence

Resolution: 0.02 Ω (0–1 Ω scale); 0.05 Ω (0–3 Ω scale)

Accuracy: ±2% of full scale

Measurement Frequency: 1000 Hz ±100 Hz

Sample Voltage: 200 µV peak at full scale; 20 mV peak maximum under all conditions

Series Reactance Tolerance: Unaffected up to 2× full-scale resistance

Superimposed DC Bias: Up to 150 V DC from external source

Applied Current: 150 / (full scale value in mΩ)

Recorder Output: 0.1 V DC at full scale; output impedance ≈1 kΩ

Power: 115/230 V ±10%, 50–60 Hz, 1.5 VA

Dimensions: 155 mm (H) × 130 mm (W) × 280 mm (D)

Weight: 3.2 kg

Key Features

Four-terminal measurement eliminates lead resistance errors

Low power dissipation enables heat-sensitive sample testing

Automatically limited voltage for safe measurement of fuses and explosive devices

150 V DC bias superposition for incremental back-biased diode resistance

Optional rechargeable battery (Option 001) supports 15 hours continuous operation

Recorder output for integration with external logging equipment

Typical Applications

Contact resistance in switches, relays, connectors

Conductor and semiconductor resistivity

Incremental diode resistance under bias

Compatibility & Integration Supplied with HP 16005A clip-type probe, HP 16006A pin-type probe, HP 16007A/B test leads, and HP 16143A probe cable. Accepts external DC bias sources up to 150 V.

Specifications

ParameterValue
Voltage / Current Range0.00001 Ω to 200 MΩ

Keysight / Agilent / HP 4328A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Keysight / Agilent / HP 4328A. Calibration covers calibration scope varies by instrument type: vacuum gauges are verified at multiple pressure points against NIST-traceable pressure standards; spectrophotometers at wavelength accuracy and photometric accuracy points; strain gauge indicators at bridge excitation accuracy and full-scale output accuracy; pH meters against certified buffer solutions at pH 4, 7, and 10.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Keysight / Agilent / HP 4328A faults: vacuum gauge filament failure (ionization gauges), spectrophotometer D2 lamp end-of-life, wavelength motor failure, strain indicator front-panel binding post corrosion, pH electrode conditioning failure. Most specialty laboratory instruments fail at their electromechanical interfaces first — vacuum gauge connectors and filament assemblies, spectrophotometer lamp housings and motor drives, strain indicator binding post corrosion. Component-level repair is standard; manufacturer support for legacy instruments is often limited, making third-party bench repair the practical path.

Keysight / Agilent / HP 4328A price

This Keysight / Agilent / HP 4328A is listed at $450. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Keysight / Agilent / HP 4328A calibration cost

NIST-traceable calibration for the Keysight / Agilent / HP 4328A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers calibration scope varies by instrument type: vacuum gauges are verified at multiple pressure points against NIST-traceable pressure standards; spectrophotometers at wavelength accuracy and photometric accuracy points; strain gauge indicators at bridge excitation accuracy and full-scale output accuracy; pH meters against certified buffer solutions at pH 4, 7, and 10.. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents