Keithley 580 Micro-ohmmeter
Keithley Laboratory Measurement Instruments. Resolution: 10 µΩ minimum, Accuracy: ±0.05% of reading + 0.1 µΩ, Current: 10 mA to 10 A at ±0.2% of reading + 0.02 mA.
- Model
- Keithley 580
- Calibration
- Available on request
- Price
- $830
- Status
- In stock
- RFQ ref
- 580
This unit
The Keithley 580 Micro-ohmmeter is a precision instrument for accurate low-resistance measurements from 200 mΩ to 200 kΩ across seven selectable ranges. It employs four-terminal Kelvin sensing to eliminate lead resistance errors, delivering 10 µΩ resolution on the lowest range and accuracy to ±0.05% of reading + 0.1 µΩ. The microprocessor-based design supports selectable pulsed or DC test current from 10 mA to 10 A, with polarity selection and automatic thermal offset cancellation in pulsed mode.
A dry circuit test function limits test voltage to 20 mV on sensitive contact junctions. Auto or manual ranging completes in 200 ms per range change, with settling time under 1 second and 3 readings per second conversion rate. The instrument displays results on a 4½ digit ±20,000 count LCD and tolerates 30V rms common mode voltage.
Optional interfaces include the Model 5802 for isolated analog output and IEEE-488 control, or the Model 1755 for digital calibration via GPIB.
Technical Specifications
- Measurement Ranges: 200 mΩ to 200 kΩ (seven ranges)
- Resolution: 10 µΩ minimum
- Accuracy: ±0.05% of reading + 0.1 µΩ
- Test Current: 10 mA to 10 A at ±0.2% of reading + 0.02 mA
- Test Voltage: 20 mV (dry circuit); 1 V (standard)
- Settling Time: <1 second to within 10 counts
- Conversion Rate: 3 readings/second typical
- Ranging: Auto (200 ms) or manual
- Display: 4½ digit LCD, ±20,000 counts
- Maximum Input Overload: 10 V limited to 10 A
- Common Mode Voltage: 30 V rms at DC, 50–60 Hz
- Temperature Coefficient: ±(0.1 × accuracy)/°C from 0–18 °C and 28–50 °C
- Power Supply: 105–125 V or 210–250 V, 50–60 Hz, 12 VA
- Dimensions: 89 mm H × 241 mm W × 300 mm D
- Weight: 3.2 kg
Key Features
- Four-terminal Kelvin sensing eliminates lead resistance errors
- Selectable pulsed or DC source current with automatic thermal offset cancellation
- Dry circuit test protects sensitive contact junctions at ≤20 mV
- Positive or negative polarity selection
- Optional battery pack (Model 1978) for ~6 hours operation
Typical Applications
- Component evaluation and material resistance characterization
- Quality control testing of low-resistance connections
- Contact resistance assessment in sensitive circuits
Compatibility & Integration
- Model 5802: Isolated analog output (1 V = 10,000 counts at ×1 gain) and IEEE-488 interface; ±(0.25% of reading + 2 mV) analog accuracy; 1000 Ω output impedance; maximum ±4 V output
- Model 1755: Digital calibration interface via IEEE-488 controller
- Rear panel banana jack connectors for measurement and external trigger inputs
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 10 mA to 10 A test current |
Keithley 580 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Keithley 580. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Keithley 580 faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.
Keithley 580 price
This Keithley 580 is listed at $830. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Keithley 580 calibration cost
NIST-traceable calibration for the Keithley 580 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →