Anrtisu 68059B Synthesized CW Generator, 0.01 – 26.5 GHz
Anrtisu 68059B Synthesized CW Generator, 0.01 – 26.5 GHz,Offers stable and accurate CW signals from 10 MHz to 26.5 GHz with low phase noise.
- Model
- Anritsu 68059B
- Calibration
- Available on request
- Price
- $17,480
- Status
- In stock
- RFQ ref
- 68059B
This unit
The Anritsu 68059B Synthesized CW Generator delivers stable, accurate continuous wave signals across 0.01 to 26.5 GHz, serving as a versatile RF/microwave stimulus source for component characterization, local oscillator replacement, and signal simulation in production and laboratory environments.
Technical Specifications
Frequency Performance- Frequency range: 0.01 to 26.5 GHz
- Resolution: 0.1 Hz (with Option 11)
- Step sweep: 1 to 10,000 steps with 1 kHz minimum step size (0.1 Hz with Option 11)
- Switching time: Less than 15 ms + 1 ms/GHz step size, or less than 40 ms maximum
- Dwell time per step: 1 ms to 99 seconds
- Manual sweep mode with user-selectable step count or step size
- Programmable frequency agility: Up to 1,000 non-sequential frequencies under GPIB control Output Characteristics
- Maximum power: +17 dBm
- Minimum power: -125 dBm
- Power resolution: 0.01 dB
- Output accuracy: 2 dB
- Output impedance: 50 Ohm
- Power level switching time: 20 ms
- Phase noise: Typically -100 dBc at 10 kHz offset (measured at 10 GHz) Signal Purity
- Harmonics: -40 dBc
- Non-harmonics: -40 dBc
Key Features
- Five sweep modes: CW, step sweep, analog sweep, alternate sweep, and manual sweep
- Alternate sweep mode enables independent sweeps between two ranges with different power levels per range
- Sweep width selectable from 1 kHz (or 0.1 Hz with Option 11) to full instrument range
- Internal modulation: AM, FM, phase modulation, pulse modulation
- User-downloadable complex modulation capability
- GPIB remote control interface
- LabWindows instrument driver available
Typical Applications
- CW stimulus generation for RF/microwave testing
- Local oscillator replacement in receiver systems
- Component analysis and characterization
- Signal simulation for ATE systems
- Phase-locked frequency stepping under automated control
Compatibility & Integration GPIB control enables integration into automated test systems and remote operation. LabWindows driver support facilitates custom application development and control.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 0.01 to 26.5 GHz | Power Range (dBm): '-125 dBm to +17 dBm |
Anritsu 68059B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu 68059B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu 68059B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Anritsu 68059B price
This Anritsu 68059B is listed at $17,480. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu 68059B calibration cost
NIST-traceable calibration for the Anritsu 68059B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →