HP / Agilent 83594A RF plug-in provides a 2 to 26.5 GHz broadband coaxial swept frequency

The HP/Agilent 83594A is an RF plug-in designed for broadband coaxial swept frequency applications, operating in the 2 to 26.5 GHz range. This module is suitable for use with HP/Agilent sweepers, providing accurate and reliable signal generation for testing and measurement purposes. It features excellent frequency accuracy and stability, making it ideal for demanding applications such as component characterization and system integration.

Model
HP / Agilent 83594A
Calibration
Available on request
Price
$1,746
Status
In stock
RFQ ref
83594A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 83594A is a broadband coaxial swept frequency RF plug-in operating from 2 to 26.5 GHz within the Agilent 8350B sweep oscillator mainframe. This source delivers continuous coverage across four internal bands with internally leveled output, making it suitable for microwave system development, receiver testing, component characterization, scalar network analysis, and local oscillator applications requiring down-conversion of high-frequency signals.

Technical Specifications Frequency Range

2 GHz to 26.5 GHz across four internal bands: 2.0–7.0 GHz, 6.9–13.5 GHz, 13.4–20.0 GHz, 19.9–26.5 GHz

Broadband switch points at approximately 2.4 GHz, 7.0 GHz, 13.5 GHz, and 20.0 GHz Power Output

Minimum output power: 10 dBm (internally leveled within 0.9 dB)

Minimum settable level: -5 dBm

Option 002 extends dynamic range to -60 dBm via 55 dB step attenuator

Display resolution: 0.1 dB

Amplitude control range: typically 15 dB

Amplitude sensitivity: 1 dB/V typical (maximum input 15 V) Output Connector

Type APC 3.5 male Slope Compensation

Calibrated range: up to 0.5 dB/GHz (10 dB over full range)

Linearity: <0.3 dB typical

Resolution: 0.1 dB/GHz Modulation

External FM: ±75 MHz (DC–100 Hz), ±7 MHz (100 Hz–1 MHz), ±5 MHz (1–2 MHz), ±1 MHz (2–10 MHz)

External AM: 100 kHz frequency response typical, ~10 kΩ input impedance Auxiliary Outputs

Fundamental oscillator output (2.0–7.0 GHz): nominally 0 dBm

Frequency reference output: nominal 1 V/GHz (2–19 GHz) via rear panel BNC Sweep Performance

Minimum single-band sweep time: 10 ms

Full-band sweep time: 25 ms Control & Integration

HP-IB programmability with 0.02 dB power level resolution

Power meter leveling compatible with HP 432A/B/C, 436A, and 438A

Key Features

Excellent harmonic performance

Completely programmable for automatic test systems

Four-band architecture optimized for broadband coverage

Power meter leveling capability

Typical Applications

Microwave system development and characterization

Receiver testing and validation

Component and subsystem characterization

Scalar network analysis

High-frequency local oscillator for down-conversion

Compatibility & Integration Designed exclusively for the Agilent 8350B sweep oscillator mainframe. Dimensions: 12 × 10 × 8 in; Weight: 10 lbs.

Specifications

ParameterValue
Frequency Range2 GHz to 26.5 GHz | Power Range (dBm): '-60 dBm to +10 dBm (with Option 002)
-5 dBm to +10 dBm (standard) | Voltage / Current Range15 V maximum input (amplitude control)

HP / Agilent 83594A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83594A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 83594A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

HP / Agilent 83594A price

This HP / Agilent 83594A is listed at $1,745.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 83594A calibration cost

NIST-traceable calibration for the HP / Agilent 83594A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents