Analogic DBS902 14-Bit Waveform Digitizer Module

Analogic VXI Modules. Resolution: 14-bit, Bandwidth: 40 MHz typical, Dynamic Range: >.

Model
Analogic DBS902
Calibration
Available on request
Price
$1,650
Status
In stock
RFQ ref
DBS902
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Analogic DBS902 is a 14-bit waveform digitizer module engineered for high-speed signal acquisition in demanding measurement environments. It functions as a plug-in component for the Analogic DBS9900 VXI C-size Dual Instrument Carrier, enabling dual-module configurations within a single chassis slot. The DBS902 delivers 80 MHz sampling with 2 MS of acquisition memory, 40 MHz typical bandwidth, and signal integrity suitable for scientific research, medical imaging, and industrial test applications.


Technical Specifications



  • Sampling Rate: Up to 80 MHz

  • Resolution: 14-bit

  • Acquisition Memory: 2 MS (MegaSamples)

  • Bandwidth: 40 MHz typical

  • Input Impedance: 1 MΩ

  • Input Capacitance: 50 pF maximum

  • Signal-to-Noise Ratio: >63 dB typical, 60 dB minimum (1 MHz Fin, 80 MS/s Fs, 40 MHz BW); >65 dB typical, 62 dB minimum (1 MHz Fin, 10 MS/s Fs, 5 MHz BW)

  • Spurious-Free Dynamic Range: >75 dBc typical, 65 dBc minimum (10 MHz Fin, 80 MS/s Fs, 40 MHz BW); >78 dBc typical, 68 dBc minimum (1 MHz Fin, 10 MS/s Fs, 5 MHz BW)

  • Input-Referred Noise: 220 µV RMS typical at 40 MHz filter, ±0.5 V range

  • Overvoltage Protection: ±3.1 VDC (low ranges), ±31 VDC (high ranges), ±124 V (DBS902HV)

  • Power Consumption: 20.0 W total (+12 VDC @ 0.50 A, -12 VDC @ 0.10 A, +5 VDC @ 2.30 A, -5.2 VDC @ 0.20 A, -2 VDC @ 0.10 A)


Key Features



  • Dual-module capability within single VXI chassis slot

  • Programmable clock sources with internal, front-panel, and VXI TTLTRG line access

  • Minimal clock skew through internal clock distribution to all plug-in modules

  • Self-triggered or externally triggered acquisition with programmable thresholds

  • Programmable sample counter for flexible data point control

  • Pre-trigger, post-trigger, and arbitrary window positioning

  • On-the-fly sample tagging via front-panel digital input for event identification and time stamping

  • VXI Plug & Play compliant with provided software driver


Compatibility & Integration Designed exclusively for the Analogic DBS9900 VXI C-size Dual Instrument Carrier. Up to two DBS902 modules install per carrier, maximizing slot utilization and reducing per-slot cost.

Specifications

ParameterValue
Frequency RangeDC to 40 MHz | Voltage / Current Range: '-12 V @ 0.10 A, -2 V @ 0.10 A, -5.2 V @ 0.20 A, +5 V @ 2.30 A, ±0.5 V to ±124 V input range

Analogic DBS902 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Analogic DBS902. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Analogic DBS902 faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.

Analogic DBS902 price

This Analogic DBS902 is listed at $1,650. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Analogic DBS902 calibration cost

NIST-traceable calibration for the Analogic DBS902 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents