Agilent / Keysight E1460A 64-Channel Relay Multiplexer
The Agilent/Keysight E1460A is a 64-channel relay multiplexer designed for signal routing and switching applications. It features high-density switching with excellent signal integrity, making it suitable for automated testing and data acquisition systems. The module offers versatile configuration options and reliable performance in demanding environments.
- Model
- Agilent / Keysight E1460A
- Calibration
- Available on request
- Price
- $2,232
- Status
- In stock
- RFQ ref
- E1460A
This unit
The Keysight E1460A is a C-size, 1-slot VXI relay multiplexer delivering high-density signal routing for automated test systems. Built around latching armature switches, this register-based module configures into multiple topologies: 64 two-wire, 32 three-wire, 32 four-wire, or 128 single-ended channels. Configuration occurs via SCPI commands or jumper settings on the terminal card, enabling flexible deployment across wire harness testing, cable validation, semiconductor characterization, and PCB diagnostics.
Technical Specifications
- Switch Architecture: Eight banks of eight Hi and Lo switches with independent commons; seven programmable control switches; six sets of wire jumpers
- Closed Channel Resistance: < 1.5 Ω (initial), < 3.5 Ω (end of life)
- Thermal Offset: 7 µV typical per channel (differential Hi-Lo)
- Maximum Voltage: 220 Vdc or 250 V rms (any terminal to any other terminal or chassis)
- Maximum Current: 1 A dc/ac rms below 30 Vdc; 0.3 A dc/ac rms below 133 Vdc (per channel common, non-inductive)
- Maximum Power: 40 VA per channel
- Insulation Resistance: 5 × 10⁸ Ω at 25 °C, 40% RH; 5 × 10⁶ Ω at 40 °C, 95% RH
- Bandwidth (–3 dB, 50 Ω source/load): 10 MHz (2-wire), 3 MHz (1-wire)
- Channel-to-Channel Crosstalk: ≤ 90 dB at 100 kHz (2-wire); ≤ 60 dB (1-wire)
- Closed Channel Capacitance: < 650 pF (Hi-Lo, 2-wire); < 700 pF (Lo-Chassis, 2-wire)
Key Features
- QUIC screw terminal block (16–26 AWG wire) with optional crimp-and-insert terminal block
- Configurable as multiple 1×8 or 1×16 two-wire multiplexers
- Register-based VXI interface (A16 slave only)
- Support for insulation testing and scanning voltmeter configurations
Typical Applications
- Wire harness and cable testing
- Semiconductor device testing
- Printed circuit board (PCB) testing
- Insulation measurement systems
- Multi-channel voltage scanning
Compatibility & Integration Connects via P1 and P2 VXI connectors. Integrates into VXI mainframes supporting register-based slave modules.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 220 V DC |
Agilent / Keysight E1460A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight E1460A. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Agilent / Keysight E1460A faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.
Agilent / Keysight E1460A price
This Agilent / Keysight E1460A is listed at $2,231.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight E1460A calibration cost
NIST-traceable calibration for the Agilent / Keysight E1460A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →