JDS Uniphase HA1503-FPS2 Optical Attenuator JDSU HA1

JDS Uniphase/JDSU Optical Attenuators. Resolution: 0.001 dB, Accuracy: ±0.1 dB.

Model
JDS Uniphase/JDSU HA1503-FPS2/HA1
Calibration
Available on request
Price
$3,430
Status
In stock
RFQ ref
HA1503-FPS2/HA1
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The JDS Uniphase HA1503-FPS2 is a programmable single-mode optical attenuator delivering ultra-high resolution attenuation control across the extended C- and L-band wavelength ranges. This benchtop instrument provides attenuation up to 100 dB with 0.001 dB resolution and ±0.1 dB accuracy, making it ideal for precision optical power characterization, signal simulation, and component testing. Front-panel FC/APC bulkhead connectors support direct integration into test setups, while GPIB and RS232 interfaces enable automated remote control via SCPI/HP8156A-compatible commands.


Technical Specifications



  • Wavelength Range: 1200 nm to 1700 nm

  • Attenuation Range: Up to 100 dB

  • Attenuation Resolution: 0.001 dB

  • Accuracy: ±0.1 dB

  • Repeatability: 0.01 dB

  • Insertion Loss: 60 dB

  • Polarization Dependent Loss (PDL): 0.03 dB typical

  • Wavelength Dependence: 90 dB infinite attenuation

  • Dual control: front-panel keypad or remote via GPIB (IEEE 488.2) and RS232

  • SCPI/HP8156A command compatibility for integration with automated test systems

  • Minimized discrete internal optical reflections and virtually eliminated cavity effects

  • Compact benchtop design with optional coupler and switch configurations


Typical Applications



  • Power meter testing and linearity calibration

  • Bit error rate (BER) and analog transmission testing

  • Loss simulation in fiber optic link characterization

  • EDFA output power analysis

  • Multichannel AM system evaluation

  • High bit-rate digital pulse code testing


Compatibility & Integration Standard configuration includes single-mode 9/125 µm fiber. Optional accessories include 10/90, 50/50, and 2/98 couplers, or a 1:2 switch for expanded test flexibility.

Specifications

ParameterValue
Wavelength Range1200 nm to 1700 nm

JDS Uniphase/JDSU HA1503-FPS2/HA1 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the JDS Uniphase/JDSU HA1503-FPS2/HA1. Calibration covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common JDS Uniphase/JDSU HA1503-FPS2/HA1 faults: motor and gear mechanism wear (attenuation steps or loss of smooth control), optical port contamination, EEPROM calibration data corruption. Most common faults are motor or gear mechanism wear on variable attenuators causing attenuation steps or loss of smooth control, optical port contamination reducing dynamic range, and EEPROM calibration data corruption on digitally controlled models.

JDS Uniphase/JDSU HA1503-FPS2/HA1 price

This JDS Uniphase/JDSU HA1503-FPS2/HA1 is listed at $3,430. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

JDS Uniphase/JDSU HA1503-FPS2/HA1 calibration cost

NIST-traceable calibration for the JDS Uniphase/JDSU HA1503-FPS2/HA1 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →