Rohde & Schwarz SMC100A 9kHz to 1.1GHz Signal Generator
Rohde & Schwarz Signal Generators. Frequency Range: 9 kHz to 1.1 GHz (expandable to 3.2 GHz with optio.
- Model
- Rohde & Schwarz SMC100A
- Calibration
- Available on request
- Price
- $4,970
- Status
- In stock
- RFQ ref
- SMC100A
This unit
The Rohde & Schwarz SMC100A is a compact analog signal generator delivering high-quality RF signal generation from 9 kHz to 1.1 GHz. Built for research, development, and production environments, it combines excellent phase noise performance, integrated modulation capabilities, and a price-to-performance ratio in a space-efficient package measuring half the width of a standard 19-inch rack and just 2 height units.
Technical Specifications
Frequency Range: 9 kHz to 1.1 GHz (expandable to 3.2 GHz with optional frequency modules) Output Power: Typical maximum output power greater than +17 dBm; +13 dBm to +17 dBm typical from 200 kHz to 3.2 GHz. Electronic attenuator with wear-free operation. Signal Quality:- SSB phase noise: Typically –111 dBc at 1 GHz (20 kHz carrier offset, 1 Hz measurement bandwidth)
- Wideband noise: Typically less than –146 dBc (f > 1 MHz, carrier offset > 10 MHz, 1 Hz BW); CW typically 10 kHz)
- Level error: Less than 0.9 dB Modulation: AM, FM, phase modulation, and pulse modulation included as standard. Maximum FM deviation 4 MHz (f > 1.6 GHz); maximum phase deviation 40 rad (f > 1.6 GHz); minimum pulse width 1 µs from integrated pulse generator. Setting Times: Frequency and level adjustments complete in less than 5 ms (typ. < 3 ms with optional IEC/IEEE bus in ALC OFF mode). Reference Frequency: 10 MHz REF OUT (+6 dBm to +12 dBm, +9 dBm typical, 50 Ω source impedance); 10 MHz REF IN (50 Ω input impedance, ±3 × 10⁻⁶ locking range, 0 dBm to +16 dBm input level range). Optional high-stability reference oscillator available.
Key Features
- Compact form factor: ½ × 19" width, 2 rack units, 3.9 kg
- Power-efficient operation suitable for lab and field deployment
- Integrated overvoltage protection
Interfaces Ethernet (TCP/IP), USB, and optional IEC/IEEE bus (R&S®SMC-K4) and GPIB connectivity.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 1.1 GHz | Power Range (dBm): '+13 dBm to +17 dBm typical (200 kHz to 3.2 GHz) |
Rohde & Schwarz SMC100A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Rohde & Schwarz SMC100A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Rohde & Schwarz SMC100A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Rohde & Schwarz SMC100A price
This Rohde & Schwarz SMC100A is listed at $4,970. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Rohde & Schwarz SMC100A calibration cost
NIST-traceable calibration for the Rohde & Schwarz SMC100A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →