Schlumberger / Solartron 1260 Impedance / Gain-Phase Analyzer
The Schlumberger / Solartron 1260 Impedance / Gain-Phase Analyzer is a precision instrument for characterizing the impedance and frequency response of electronic components, circuits, and materials. It delivers accurate measurements over a wide frequency range. Ideal for research, development, and quality control applications.
- Model
- Schlumberger / Solartron 1260
- Calibration
- Available on request
- Price
- $310
- Status
- In stock
- RFQ ref
- 1260
This unit
The Schlumberger/Solartron 1260 is a precision impedance and gain-phase analyzer engineered for accurate characterization of frequency response and impedance across materials and electronic components. It employs a patented single-sine correlation technique to reject noise and harmonic distortion, delivering 0.1% impedance accuracy and 0.1° phase accuracy. The instrument measures impedance from 100 mΩ to >100 MΩ (extendable to 100 TΩ with the 1296A Dielectric Interface) and supports 2-, 3-, and 4-terminal measurement configurations.
It operates over 10 µHz to 32 MHz with 0.015 ppm frequency resolution, enabling broad applicability in materials research, electrochemistry, and component development.
Technical Specifications
- Frequency Range: 10 µHz to 32 MHz
- Frequency Resolution: 0.015 ppm (1 in 65 million)
- Impedance Measurement Range: 100 mΩ to >100 MΩ; up to 100 TΩ with 1296A Dielectric Interface
- Impedance Accuracy: 0.1%
- Phase Accuracy: 0.1°
- Polarization Voltage: Up to ±40.95 V
- Measurement Configurations: 2-, 3-, and 4-terminal
Key Features
- Patented single-sine correlation technique for inherent noise and harmonic distortion rejection
- Measurement modes: gain, phase, group delay, impedance, inductance, capacitance, quality factor, dissipation factor, voltage and current transfer characteristics, and immittance
- Frequency, amplitude, and bias sweep capability
- Null, normalize, and auto-impedance modes
- Built-in statistical analysis, limit checking, and data reduction
- Component sorting (manual or automatic)
- Self-test and learn program facilities
- Stray capacitance and inductance compensation via result scaling
- GPIB and RS 423 interfaces; plotter output; history file export
- Standalone or computer-controlled operation with SMaRT software and optional ZPLOT
Typical Applications
- Corrosion studies
- Battery research and fuel cells
- Solar cells and LCDs
- Bio-materials and dielectric analysis
- Ceramics and composites
- Electronic component development
- Electrochemical measurements
- Materials analysis and civil engineering
Compatibility & Integration The 1260 integrates with Solartron Analytical products including the 1296A Dielectric Interface for extended impedance range and the 1287A Electrochemical Interface.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 µHz to 32 MHz | Voltage / Current Range: ±40.95 V |
Schlumberger / Solartron 1260 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Schlumberger / Solartron 1260. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Schlumberger / Solartron 1260 faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.
Schlumberger / Solartron 1260 price
This Schlumberger / Solartron 1260 is listed at $310. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Schlumberger / Solartron 1260 calibration cost
NIST-traceable calibration for the Schlumberger / Solartron 1260 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →