Astronics / EADS / Racal 1260-30C Multiplexer VXI Module

The Astronics/EADS/Racal 1260-30C is a Multiplexer VXI Module designed for various data acquisition and signal routing applications. This module is ideal for automated test equipment (ATE) systems, data logging, and process monitoring. It offers high-speed switching and reliable performance within a VXI environment.

Model
Astronics / EADS / Racal 1260-30C
Calibration
Available on request
Price
$1,150
Status
In stock
RFQ ref
1260-30C
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Astronics/EADS/Racal 1260-30C is a high-power multiplexer VXI module engineered for signal routing and switching in automated test equipment systems. This C-size, single-slot VXIbus module delivers flexible channel configurations with switching ratings up to 250V and 2A per channel, making it suitable for both DC and AC test applications across laboratory and industrial environments.

Technical Specifications Switching Capacity & Configurations

Maximum voltage: 220V DC or 250V AC RMS (terminal-terminal or terminal-chassis)

Maximum current: 2A per channel

Maximum power: 60W DC or 62.5V A per channel

Quad 1×10 multiplexer architecture configurable as one 1×40, two 1×20, four 1×10, or eight 1×5 two-wire network DC Performance

Path resistance: 10⁹ Ω AC Performance (50Ω load)

Bandwidth (−3 dB): Typical 35 MHz

Open channel capacitance: <25 pF

Channel-chassis capacitance: <25 pF

High-low capacitance: <50 pF

Insertion loss: <0.2 dB @ 100 kHz, <0.25 dB @ 1 MHz, <0.5 dB @ 10 MHz

Crosstalk: <−75 dB @ 100 kHz, <−60 dB @ 1 MHz, <−40 dB @ 10 MHz Power Requirements

+5V: 1A nominal (0.4A without Option 01; 2.8A with Option 01)

+24V: 10 mA per energized relay Physical & Environmental

VXIbus C-size form factor, single-slot module

Operating temperature: 0°C to +55°C

Non-operating temperature: −40°C to +71°C

Relative humidity: 95% ±5% non-condensing at 30°C

Key Features

Relay coil current monitoring for confidence checking

User-supplied component mounting holes and terminator circuit pads

CE Mark compliance for RF emissions, immunity, and electrical safety standards

Message-based and register-based control via Option 01T interface (Option 01 Smart Card required)

Typical Applications Data logging, process monitoring, and multi-channel signal switching in automated test environments requiring high-voltage and high-current switching with low crosstalk.

Compatibility & Integration VXIbus C-size, single-slot module integrating directly into VXI-based ATE systems via message and register-based control protocols.

Specifications

ParameterValue
Frequency RangeDC to 35 MHz | Voltage / Current Range: 0-250V AC RMS or 220V DC, 0-2A per channel

Astronics / EADS / Racal 1260-30C — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Astronics / EADS / Racal 1260-30C. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Astronics / EADS / Racal 1260-30C faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.

Astronics / EADS / Racal 1260-30C price

This Astronics / EADS / Racal 1260-30C is listed at $1,150. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Astronics / EADS / Racal 1260-30C calibration cost

NIST-traceable calibration for the Astronics / EADS / Racal 1260-30C is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →