Fluke 6082A – Synthesized RF Signal Generator

The Fluke 6082A is a synthesized RF signal generator that delivers high-performance and reliability for a wide range of RF testing applications. It offers precise frequency control, versatile modulation capabilities, and a clean spectral output. Ideal for communication, aerospace, and defense industries.

Model
Fluke 6082A
Calibration
Available on request
Price
$3,400
Status
In stock
RFQ ref
6082A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Fluke 6082A is a synthesized RF signal generator engineered for RF testing across communication, aerospace, and defense applications. Operating from 100 kHz to 2112 MHz with 0.001 Hz resolution, it delivers output power from -140 dBm to +19 dBm (with overranging to +20 dBm) into a 50 Ω nominal impedance. The instrument combines exceptional spectral purity - phase noise of -131 dBc/Hz at 1 GHz (20 kHz offset) and non-harmonic spurious below -100 dBc - with versatile modulation. AM, FM, pulse, and phase modulation operate independently or combined, with FM deviation to 8 MHz across rates from 0 to 100 kHz. Pulse rise time is <15 ns typical, and the internal modulation oscillator provides coherence with the RF carrier. Linear and logarithmic frequency or amplitude sweeps operate over 1 ms to 1000 s with digital control. Output level accuracy ranges from ±1 dB (0.4–2112 MHz) to ±2 dB (0.1–0.4 MHz), with ±1 dB flatness at +10 dBm. A fixed 20 dB attenuator vernier delivers transient-free switching in 0.1 dB steps. The instrument supports FSK and ASK modulation types and includes IEEE-488 (GPIB) control with emulation modes for Fluke 6060/6070 compatibility.

Technical Specifications

Frequency Range: 100 kHz to 2112 MHz, 0.001 Hz resolution

Output Power: -140 dBm to +19 dBm (±1 dB accuracy, 0.4–2112 MHz); ±1 dB flatness

Output Impedance: 50 Ω nominal

Phase Noise: -131 dBc/Hz at 1 GHz (20 kHz offset)

Non-Harmonic Spurious: <-100 dBc below 1056 MHz; <-94 dBc above 1056 MHz

FM Deviation: Up to 8 MHz; FM rates 0–100 kHz

Pulse Rise Time: <15 ns typical (<7.5 ns typical); 80 dB on/off ratio

Modulation Bandwidth: 500 kHz

Sweep Range: 1 ms to 1000 s; linear or logarithmic

Attenuator: 20 dB vernier, 0.1 dB resolution, transient-free

Control Interface: IEEE-488 (GPIB)

Key Features

Combined AM, FM, pulse, and phase modulation for complex signal synthesis

Coherent internal modulation oscillator enabling RF-carrier-coherent pulse waveforms

FSK and ASK capability

Digital frequency and amplitude sweep with 4096 maximum steps

X-axis (0 to +10 V ±10%) and Z-axis (TTL) sweep outputs

Low 1.5 Hz residual FM

Typical Applications

RF receiver and transceiver testing

Modulation performance validation

Spectral purity and spurious measurement

Aerospace and defense signal simulation

Compatibility & Integration IEEE-488 (GPIB) standard interface with Fluke 6060/6070 software emulation modes.

Specifications

ParameterValue
Frequency Range100 kHz to 2112 MHz | Power Range (dBm): '-140 dBm to +19 dBm | Voltage / Current Range: 0 to +10 V ±10% (X-axis output)

Fluke 6082A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Fluke 6082A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Fluke 6082A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Fluke 6082A price

This Fluke 6082A is listed at $3,400. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Fluke 6082A calibration cost

NIST-traceable calibration for the Fluke 6082A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →