Anritsu 68397B Synthesized Sweep/Signal Generator, 0.01 – 65 GHz
Anritsu 68397B,"Synthesized sweep/signal generator covering 0.01 to 65 GHz for wideband RF testing. Offers precise frequency control, low phase noise, and stable output. Designed for testing high-frequency components and systems. Suitable for telecom, aerospace, and microwave applications requiring accurate signal simulation across broad frequency ranges."
- Model
- Anritsu 68397B
- Calibration
- Available on request
- Price
- $32,950
- Status
- In stock
- RFQ ref
- 68397B
This unit
The Anritsu 68397B is a synthesized sweep/signal generator delivering precise RF signal generation across 0.01 GHz to 65 GHz. Built for telecommunications, aerospace, defense, and microwave R&D, it combines low phase noise, stable output power, and flexible sweep capabilities to characterize components, sub-assemblies, and complete systems. The instrument supports frequency and amplitude sweeps, step and list modes, with user-selectable sweep times from 1 ms to 1000 s and triggering options including internal, external, manual, and line.
Technical Specifications
Frequency Performance- Frequency range: 0.01 GHz to 65 GHz
- Resolution: 0.01 Hz with user-selectable steps down to 1 mHz
- Switching speed: <5 ms between any two frequencies
- Setting time: <10 ms for frequency and amplitude Output Characteristics
- Power range: -120 dBm to +10 dBm (varies with frequency)
- Power accuracy: ±1.0 dB at +10 dBm, 1 GHz
- Power flatness: ±0.5 dB across 1 GHz bandwidth at +10 dBm
- Output impedance: 50 Ohms nominal
- VSWR: <2.0:1 from 1 GHz to 65 GHz Signal Purity
- Phase noise: <-110 dBc/Hz at 10 kHz offset (10 GHz); <-100 dBc/Hz at 10 kHz offset (60 GHz)
- Harmonic distortion: <-35 dBc at 0 dBm output, <40 GHz
- Spurious outputs: <-60 dBc at <40 GHz, 70 dB on/off ratio, 10 ns to CW pulse width, 10 Hz to 100 kHz PRF
Key Features
- Four sweep modes: frequency, amplitude, step, and list
- Multiple triggering options: internal, external, manual, line
- Connectors: Type-N (female) to 18 GHz; 2.4 mm (female) above 18 GHz
Compatibility & Integration IEEE-488.2 (GPIB), Ethernet (10/100 Base-T), and USB control with SCPI command language support enable laboratory and production-line integration.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 0.01 GHz to 65 GHz | Power Range (dBm): '-120 dBm to +10 dBm |
Anritsu 68397B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu 68397B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu 68397B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Anritsu 68397B price
This Anritsu 68397B is listed at $32,950. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu 68397B calibration cost
NIST-traceable calibration for the Anritsu 68397B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →