Fluke 8060A True RMS Digital Multimeter
The Fluke 8060A is a handheld true RMS digital multimeter designed for electronics troubleshooting and general purpose measurements. It features a 4.5-digit LCD display, true RMS AC voltage and current measurements, resistance, continuity, diode test, and frequency measurement capabilities. Ideal for field service, benchtop testing, and electrical maintenance.
- Model
- Fluke 8060A
- Calibration
- Available on request
- Price
- $110
- Status
- In stock
- RFQ ref
- 8060A
This unit
The Fluke 8060A is a microcomputer-based 4.5-digit handheld digital multimeter delivering True RMS measurement capabilities for accurate AC voltage and current readings on non-sinusoidal waveforms. The instrument combines precision and reliability across voltage, current, resistance, frequency, conductance, continuity, diode, and decibel measurements - making it essential for engineers and technicians in audio, video, telecommunications, and computer technology.
Technical Specifications Display and Resolution
4.5-digit LCD with 20,000 counts. DC Voltage (VDC) Ranges: 200.00 mV, 2.0000 V, 20.000 V, 200.00 V, 1000.0 V. Best accuracy ±(0.04% + 2). Input impedance >1000 MΩ. **AC Voltage (VAC)
True RMS, AC Coupled** Ranges: 200.00 mV, 2.0000 V, 20.000 V, 200.00 V, 750.0 V. Accuracy ±(0.2% + 10). Frequency response to 100 kHz. DC Current (ADC) Ranges: 200.00 µA, 2.0000 mA, 20.000 mA, 200.00 mA, 2000.0 mA. Best accuracy ±(0.2% + 2). **AC Current (AAC)
True RMS** Ranges: 200.00 µA, 2.0000 mA, 20.000 mA, 200.00 mA, 2000.0 mA. Accuracy ±(0.75% + 10). Frequency response to 30 kHz. Resistance Ranges: 200.00 Ω to 300.0 MΩ. Best accuracy ±(0.07% + 2). Includes low-power ohms for in-circuit testing. Frequency Ranges: 200.00 Hz, 2.0000 kHz, 20.000 kHz, 200.00 kHz. Accuracy ±(0.05% + 1). Measurement span 12 Hz to over 200 kHz. Conductance Range: 2000.0 nS. Accuracy ±(0.5% + 20). Continuity Fast audible and visual indication. Recognizes continuity as brief as 50 microseconds. Diode Test Constant-current mode. Range 2 V, test current 1 mA. Decibels (dBm) Reference: 600 Ω. Ranges: -50.00 to +60.00 dBm. Accuracy ±(0.15 dB).
Key Features
True RMS measurement for non-sinusoidal waveforms
Relative reference (REL) function stores reference values to display differences across voltage, current, resistance, conductance, frequency, and decibel measurements
Software-controlled self-test for internal circuit verification
Double-fused current inputs with extensive overload protection
Single 9V battery with typical operating life of 170 hours
Typical Applications Audio and video system testing, telecommunications equipment troubleshooting, computer hardware diagnostics, and general electronic circuit characterization.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 Hz to 100 kHz | Voltage / Current Range: DC Voltage: 600.0 mV to 1000 V |
| AC Voltage | 600.0 mV to 1000 V |
| DC Current | 60.00 mA to 10.00 A |
| AC Current | 60.00 mA to 10.00 A |
Fluke 8060A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Fluke 8060A. Calibration covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Fluke 8060A faults: blown input fuse, worn input jacks, display segment failure, relay contact wear on bench DMMs. Common faults include worn input jacks (banana socket contact wear on the mA and COM terminals), blown input fuses (especially the 440 mA HRC fuse), and display segment failures. Keithley 2000/2001/2002 bench DMMs commonly develop relay contact wear affecting resistance accuracy after extended service.
Fluke 8060A price
This Fluke 8060A is listed at $110. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Fluke 8060A calibration cost
NIST-traceable calibration for the Fluke 8060A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. Turnaround is 1 to 3 business days after receipt. See full calibration details →