HP / Agilent 83631A Synthesized Source, 45 MHz to 26.5 GHz

The HP/Agilent 83631A is a synthesized signal generator that delivers high performance and reliability. It offers wide frequency coverage from 45 MHz to 26.5 GHz, excellent signal quality, and precise control over output parameters. Ideal for demanding applications in R&D, manufacturing, and service environments.

Model
HP / Agilent 83631A
Calibration
Available on request
Price
$11,158
Status
In stock
RFQ ref
83631A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 83631A is a synthesized signal source engineered for precise microwave signal generation across 45 MHz to 26.5 GHz. Purpose-built for integration with Agilent 8510 network analyzer systems, this instrument combines the stability of a synthesized source with sweep oscillator versatility. It delivers 1 Hz frequency resolution and operates without modulation capabilities or front-panel controls - all configuration and output management occurs via rear-panel connections and programming interfaces.


Technical Specifications



  • Frequency Range: 45 MHz to 26.5 GHz

  • Frequency Resolution: 1 Hz

  • CW Mode Accuracy: Same as time base accuracy

  • Swept Mode Accuracy:


Sweep widths ≤ n × 10 MHz: 0.1% of sweep width ± time base accuracy


Sweep widths > n × 10 MHz and ≤ 400 MHz: 1% of sweep width


Sweep widths > 400 MHz and ≤ 4 GHz: 4 MHz



  • Minimum Settable Output Power: −20 dBm

  • Dimensions: 24 × 20 × 16 in

  • Weight: 50 lbs


Key Features



  • No modulation circuitry - pure, unmodulated signal generation

  • Rear-panel output connector with no front-panel keyboard or display

  • Programmable via network analyzer language, SCPI, or M.A.T.E. CIIL (Option 700)

  • Rear-panel switch selects power-up language configuration

  • Rack-mountable with optional slide kit (806) or flange kits (908, 913)


Typical Applications



  • Receiver characterization and testing

  • Radar system development

  • Wireless communication signal sourcing


Compatibility & Integration Optimized for Agilent 8510C network analyzer systems. Compatible with HP 8970B noise figure meter when configured to the appropriate HP-IB address and network analyzer language. Not compatible with HP 8757C/E systems.

Specifications

ParameterValue
Frequency Range45 MHz to 26.5 GHz | Power Range (dBm): −20 dBm (minimum settable)

HP / Agilent 83631A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83631A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 83631A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

HP / Agilent 83631A price

This HP / Agilent 83631A is listed at $11,158. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 83631A calibration cost

NIST-traceable calibration for the HP / Agilent 83631A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →