HP / Agilent 8443B TRACKING GEN

HP / Agilent RF Signal Sources. Accuracy: ** ±0.3 dB (absolute at 0 dBm, Resolution: ** <, Accuracy: ** <.

Model
HP / Agilent 8443B
Calibration
Available on request
Price
$628
Status
In stock
RFQ ref
8443B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 8443B Tracking Generator is a dual-function instrument that generates swept RF signals synchronized with spectrum analyzer tuning while providing integrated frequency measurement capabilities. The unit delivers tracking generator output from 100 kHz to 110 MHz at 50 Ohms impedance, with amplitude control spanning -120 dBm to +10 dBm. Front-panel controls offer 10 dB and 1 dB stepped attenuation plus continuous vernier adjustment for precise level setting. Amplitude flatness remains within ±0.5 dB across the full frequency range, with absolute accuracy of ±0.3 dB at 0 dBm and 30 MHz.

Technical Specifications

RF Output: 100 kHz to 110 MHz, 50 Ohms, AC coupled via BNC connector

Output Level: -120 dBm to +10 dBm with ±0.5 dB flatness

Amplitude Accuracy: ±0.3 dB (absolute at 0 dBm, 30 MHz); attenuator: 10 dB steps ±0.2 dB, 1 dB steps ±0.1 dB; vernier: 1.2 dB continuous

Frequency Resolution: <10 Hz

Time Base Accuracy: <3 × 10⁻⁹ per day after warmup (0.3 Hz/day)

Time Base Temperature Drift: <3 × 10⁻⁸ (3 Hz) over 0–55°C range

Power: 115 V or 230 V, 48–440 Hz, 75 W active / 30 W standby

Operating Temperature: 0 to 55°C; Storage: -40 to +75°C

Dimensions: 16 × 14 × 10 inches; Weight: ~25 lbs

Key Features

Integrated 7-digit frequency counter (plus 1 overrange digit) covering 100 kHz to 110 MHz

Counter resolution selectable at 1 kHz (1 mS gate), 100 Hz (10 mS gate), or 10 Hz (100 mS gate)

Counter accuracy: ±1 count ± time base accuracy

Three counter modes: Track Analyzer Mode, Restore Mode, and Scan Hold Mode

External counter input: 10 kHz to 120 MHz, 50 Ohms, -10 to +25 dBm

Digital frequency output in 8,4,2,1 code format

Time base input: 1 MHz, 40 Ohms, 1 Vrms minimum

Typical Applications

Frequency response characterization of RF filters and amplifiers

Insertion loss and gain measurements

Component alignment and alignment verification

Swept impedance analysis in conjunction with compatible spectrum analyzers

Compatibility & Integration Designed for use with HP/Agilent spectrum analyzers such as the 8553B, providing automatic frequency tracking during swept measurements at analyzer marker positions.

Specifications

ParameterValue
Frequency Range100 kHz to 110 MHz | Power Range (dBm): '-120 dBm to +10 dBm | Voltage / Current Range: 115 V or 230 V, 48-440 Hz, 75 W active

HP / Agilent 8443B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 8443B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 8443B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

HP / Agilent 8443B price

This HP / Agilent 8443B is listed at $628. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 8443B calibration cost

NIST-traceable calibration for the HP / Agilent 8443B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents