Agilent / Keysight E1343A 16-Channel GP High-Voltage Relay Multiplexer
Agilent / Keysight VXI Modules. Channel: 1 VA.
- Model
- Agilent / Keysight E1343A
- Calibration
- Available on request
- Price
- $2,410
- Status
- In stock
- RFQ ref
- E1343A
This unit
The Agilent/Keysight E1343A is a 1-slot B-size VXI relay multiplexer module delivering high-voltage switching across 16 channels with configurable architectures. Built on reed relay technology, the module supports operation as a single 1×16 three-wire multiplexer, two independent 1×8 three-wire multiplexers, or a 1×8 six-wire multiplexer - enabling flexible test configurations from simple analog routing to four-wire resistance measurements and precision 6-wire setups. Register-based control via SCPI commands and downloadable firmware provide straightforward integration into automated test systems.
Technical Specifications
Switching Performance- 16 channels, each with high, low, and guard connections
- Reed relay construction with break-before-make switching
- 350 channels/second typical scanning rate
- Relay life: 10⁸ operations (no load); 10⁷ operations (rated load)
- Relays open on power-up and power-down Electrical Ratings
- Maximum voltage (any terminal to any other or chassis): 250 Vdc / 250 Vrms ac
- Maximum current per channel common (non-inductive): 50 mA
- Maximum power per channel: 1 VA Module Architecture
- VXI device type: Register-based, A16, slave only
- Form factor: B-size, 1 slot
- Connector: P1
- Screw terminal connections: 16–26 AWG wire (0.5–1.5 mm)
- Weight: 10 lbs (approximate)
- Dimensions: 12 × 10 × 8 in (approximate) Power Consumption
- +5 V: 0.2 A
- +12 V: 0.13 A
- +24 V: 0 A
- −5.2 V: 0 A
- −12 V: -
Key Features
- Multiple configuration modes for general-purpose analog switching
- Shared switch card design with E1345A, E1346A, E1347A, E1355A, E1356A modules
- Analog bus connectivity for inter-module communication
- SCPI-based control with downloadable command module firmware
Compatibility & Integration The E1343A fits directly into B-size VXI mainframes (Agilent E1300B, E1301B). For C-size mainframe integration, deploy the E1403C active adapter. Analog bus cables enable connection with companion instruments such as the E1326B DMM. Instrument drivers and software support include I-SCPI (Win 3.1, Series 700), C-SCPI (LynxOS, Series 700), VXIplug&play frameworks, and panel drivers.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 250 Vdc maximum |