Agilent/Keysight 1332A
Agilent / Keysight VXI Modules. Bandwidth: DC to 5 MHz at 7.6 cm (3") deflection, Channels: 4 channels (expandable to 8 via multiplexing: 4 is.
- Model
- Agilent / Keysight 1332A
- Calibration
- Available on request
- Price
- $97
- Status
- In stock
- RFQ ref
- 1332A
This unit
The Agilent/Keysight 1332A is a modular display unit and VXI counter instrument designed for precision signal analysis in electronic test environments. When configured as an X-Y display, it delivers fast signal settling and high writing speed for real-time waveform visualization. As a VXI module, it functions as a 4-channel counter/totalizer supporting frequency measurement, period averaging, pulse width analysis, time interval measurement, totalization, gated totalization, and up/down counting.
The instrument integrates into larger VXI ecosystems for multi-function measurement automation.
Technical Specifications
Display Characteristics (X-Y Display Configuration):- Bandwidth: DC to 5 MHz at 7.6 cm (3") deflection
- Rise time: ≤ 70 ns (10% to 90% points)
- Phase shift: 25.4 cm/µs (10 in./µs)
- Input impedance: High impedance via rear panel BNC connectors (shield grounded)
- Maximum input: ±50 V (DC + peak AC)
- Differential inputs available (optional) VXI Module Specifications (E1332A 4-Channel Counter/Totalizer):
- Module type: B-size, 1-slot, register-based VXI device
- Channels: 4 channels (expandable to 8 via multiplexing: 4 isolated, 4 non-isolated)
- Simultaneous totalization: 4 channels
- Frequency measurement: Up to 4 MHz (channels 1 and 3); resolution in Hz or gate time 2 ms to 65.5 s in 16 binary steps; minimum pulse width 125 ns
- Period average: Channels 1 and 3; averages 2^N periods; N range 1 to 16; resolution 1 / (5 × 10^6 × 2^N) s; minimum pulse width 125 ns
- Pulse width measurement: Channels 2 and 4; range 500 ns to 858 s; resolution 200 ns; positive or negative edge selectable
- Time interval: Between channel transitions; selectable rising or falling edge; minimum interval 500 ns
Key Features
- Sub-100 ns rise time enables transient capture with minimal distortion
- Seven distinct counter functions in single 1-slot module
- Multiplexed channel expansion for extended measurement capacity
- Adjustable deflection factors accommodate varied signal levels
- Low phase shift maintains measurement accuracy across DC and AC spectra
Typical Applications
- Real-time oscillographic signal display and analysis
- Multi-channel frequency and time interval measurements
- Pulse characterization and timing analysis
- Automated test equipment (ATE) integration via VXI backplane
- Gated counting and statistical accumulation
Compatibility & Integration The 1332A operates within VXI modular instrumentation frameworks. Display and counter configurations can operate independently or as complementary components in larger measurement systems.