Fluke 83V Industrial Multimeter, 6000 counts

The Fluke 83V Industrial Multimeter is a solid and reliable tool for troubleshooting and diagnostics in industrial settings. It features a 6000-count display for accurate readings, measures AC and DC voltage, current, resistance, capacitance, and frequency. It also includes a built-in thermometer for temperature measurements and Min/Max/Average recording to capture intermittent signals.

Model
Fluke 83V
Calibration
Available on request
Price
$349
Status
In stock
RFQ ref
83V
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Fluke 83V Industrial Multimeter delivers precision electrical measurement across demanding industrial environments. Its 6000-count digital display updates 4/sec, paired with a 32-segment analog bargraph refreshing at 40/sec, enabling technicians to diagnose faults in control rooms, plant floors, and high-altitude installations. The instrument combines broad measurement capability - from microvolts to kilovolts, microamps to 20 A impulse - with safety features designed to withstand transient overvoltages exceeding 8,000 V.

Technical Specifications Voltage & Current

DC voltage: 600.0 mV to 1000 V with ±(0.1% + 1) accuracy; 100 µV resolution

AC voltage: 600.0 mV to 1000 V with ±(0.5% + 2) accuracy; 0.1 mV resolution (average responding, RMS indicating)

DC current: 600.0 µA to 10.00 A (20 A for up to 30 seconds); ±(0.4% + 2) accuracy; 0.01 mA resolution

AC current: 600.0 µA to 10.00 A (20 A for up to 30 seconds); ±(1.2% + 2) accuracy; 0.1 µA resolution Passive Components & Frequency

Resistance: 50 MΩ maximum; ±(0.4% + 1) accuracy; 0.1 Ω resolution

Capacitance: 9,999 µF maximum; ±(1% + 2) accuracy; 0.01 nF resolution

Frequency: 200 kHz maximum; ±(0.005% + 1) accuracy; 0.01 Hz resolution (frequency display: 19,999 counts, updates 3/sec above 10 Hz)

Conductance: 60.00 nS maximum; ±(1.0% + 10) accuracy; 0.01 nS resolution

Duty cycle: 99.9% maximum; ±(0.2% per kHz + 0.1%) accuracy; 0.1% resolution

Diode test: 3 V range; ±(2% + 1) accuracy; 1 mV resolution

Key Features

Auto and manual ranging for adaptive measurement

Min/Max/Average recording with alert capability for transient event capture

Relative mode for test lead resistance elimination in low-ohm circuits

Touch Hold function for noise-free stable reading acquisition

Input Alert audible warning against jack misconnection

Bright white, two-level backlit display and keypad for high-visibility operation

Selectable Sleep Mode for extended battery operation

EN 61010-1 compliance: CAT III to 1000 V, CAT IV to 600 V

Impulse withstand rating exceeding 8,000 V for surge protection

Typical Applications Control room diagnostics, plant floor troubleshooting, high-altitude field service, industrial circuit analysis, preventive maintenance testing.

Specifications

ParameterValue
Frequency RangeDC to 200 kHz | Voltage / Current Range: 600.0 mV to 1000 V DC/AC, 600.0 µA to 10.00 A DC/AC (20 A for up to 30 seconds)

Fluke 83V — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Fluke 83V. Calibration covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Fluke 83V faults: blown input fuse, worn input jacks, display segment failure, relay contact wear on bench DMMs. Common faults include worn input jacks (banana socket contact wear on the mA and COM terminals), blown input fuses (especially the 440 mA HRC fuse), and display segment failures. Keithley 2000/2001/2002 bench DMMs commonly develop relay contact wear affecting resistance accuracy after extended service.

Fluke 83V price

This Fluke 83V is listed at $349. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Fluke 83V calibration cost

NIST-traceable calibration for the Fluke 83V is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. Turnaround is 1 to 3 business days after receipt. See full calibration details →