Exfo IQ-5320 Multi-Wavelength Meter
Exfo DWDM / CWDM Test Equipment. Resolution: ** 0.001 nm, Power Range: ** +10 dBm to −45 dBm, Resolution: ** 0.01 dB.
- Model
- Exfo IQ5320
- Calibration
- Available on request
- Price
- $3,200
- RFQ ref
- IQ5320
This unit
The EXFO IQ-5320 is a precision multi-wavelength meter engineered for simultaneous measurement of central wavelength and optical power across up to 200 multiplexed channels. Built on a scanning Michelson interferometer architecture, it delivers wavelength accuracy superior to traditional optical spectrum analyzers - ideal for DWDM channel verification and power mapping in production and field test environments.
Technical Specifications
Wavelength Range: 1450 nm to 1650 nm
Wavelength Uncertainty: ±0.003 nm absolute; ±0.002 nm typical; ±0.005 nm maximum
Wavelength Display Resolution: 0.001 nm
Wavelength Repeatability: ±0.001 nm
Wavelength Measurement Resolution: 0.001 nm
Channel Discrimination: <0.06 nm (similar power); <0.2 nm (dissimilar power)
Optical Power Range: +10 dBm to −45 dBm
Power Display Resolution: 0.01 dB
Optical Power Accuracy: ±0.35 dB; ≤±0.1 dB at 25°C, 1550 nm, −10 dBm
Optical Power Repeatability: ±0.1 dB
Optical Power Flatness: <0.2 dB
Optical Power Linearity: <0.1 dB (−20 to +10 dBm); 23 dB
Return Loss: >40 dB (APC connector)
Polarization Dependence: <0.4 dB
Simultaneous Channels: Up to 200
Physical Size: 12.5 cm H × 7.4 cm W × 28.2 cm D; 15 lbs
Key Features
Spectrum view window with Fabry-Perot modal distribution display
Zoom and marker functions for detailed spectral analysis
Average acquisition mode for improved signal-to-noise performance
Drift mode for long-term wavelength and power variation tracking
Raw spectrum data output for proprietary algorithm integration
System Integration Operates as a PC-based module within EXFO's IQS-500 Intelligent Test System. Data acquisition via GPIB standard interface. LabVIEW drivers and ActiveX libraries support custom applications in networked PC environments. Backward-compatible with most IQ-generation modules.
Typical Applications DWDM system verification, optical channel power mapping, production test automation, wavelength and power characterization.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '+10 dBm to −45 dBm | Wavelength Range: 1450 nm to 1650 nm |
Exfo IQ5320 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Exfo IQ5320. Calibration covers absolute wavelength accuracy at multiple reference points across the full measurement range, frequency accuracy, wavenumber accuracy, and multi-wavelength channel spacing measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Exfo IQ5320 faults: interferometer alignment drift, reference etalon or HeNe laser end-of-life, InGaAs detector sensitivity loss, fiber input connector damage. Fizeau interferometer-based instruments (Agilent 86120B/86122A/C) most commonly develop interferometer alignment drift and InGaAs detector degradation over time. Burleigh WA-series scanning Fabry-Perot instruments most commonly fail due to reference HeNe laser end-of-life. Both types are fully repairable with post-repair wavelength accuracy verification.
Exfo IQ5320 price
This Exfo IQ5320 is listed at $3,200. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Exfo IQ5320 calibration cost
NIST-traceable calibration for the Exfo IQ5320 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers absolute wavelength accuracy at multiple reference points across the full measurement range, frequency accuracy, wavenumber accuracy, and multi-wavelength channel spacing measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →