Rohde & Schwarz SMR40 10MHz – 40GHz Signal Generator

Rohde & Schwarz RF Signal Sources. Frequency Range: 10 MHz to 40 GHz, Resolution: 0.1 Hz.

Model
Rohde & Schwarz SMR40
Calibration
Available on request
Price
$6,400
Status
In stock
RFQ ref
SMR40
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Rohde & Schwarz SMR40 is a microwave signal generator spanning 10 MHz to 40 GHz, engineered for receiver measurements, scalar network analysis, and demanding RF test scenarios across aerospace, defense, and communications. It delivers exceptional spectral purity through advanced fractional-N frequency synthesis, with output power exceeding +9 dBm at 40 GHz and adjustable levels down to -130 dBm via optional RF attenuation. The instrument combines low SSB phase noise (< -83 dBc at 10 GHz, 10 kHz offset) with excellent spurious suppression and harmonic rejection - below -55 dBc from 30 MHz to 20 GHz, and below -40 dBc above 20 GHz.


Technical Specifications



  • Frequency Range: 10 MHz to 40 GHz

  • Frequency Resolution: 0.1 Hz

  • Output Power: Greater than +9 dBm at 40 GHz; adjustable with compensation above -20 dBm

  • Output Range (with optional attenuator): -130 dBm to +9 dBm

  • SSB Phase Noise: < -83 dBc (10 GHz, 10 kHz offset, CW, FM off, 1 Hz bandwidth)

  • Harmonics Suppression: < -55 dBc (30 MHz < f ≤ 20 GHz); 20 GHz)

  • Non-Harmonics: < -60 dBc (f ≤ 20 GHz); 20 GHz)


Key Features



  • Pulse Modulation (Standard): Internal pulse generator with on/off ratio better than 80 dB; rise/fall times shorter than 12 ns; widths to 25 ns

  • Optional Modulation (B5): AM/FM/LF generator (0.1 Hz to 10 MHz), FM bandwidth DC to 5 MHz, FSK data rates 0 to 2 MHz, ASK, scan, and phase modulation; all modes combinable

  • Sweep Capability (Standard): Digital RF, level, and frequency sweep; 10 user-selectable frequency markers

  • Optional Analog Sweep (B4): Synthesized ramp sweep to 600 MHz/ms (f > 2 GHz)

  • Operating Modes: Automatic, singleshot, manual, externally triggered

  • Calibration Cycle: 3 years


Typical Applications Receiver characterization, network scalar analysis, component testing, and system validation in microwave and millimeter-wave development and production environments.


Compatibility & Integration Highly configurable with optional modules including OCXO reference oscillator (Option B1) for enhanced frequency stability and aging performance.

Specifications

ParameterValue
Frequency Range10 MHz to 40 GHz | Power Range (dBm): '-130 dBm to +9 dBm

Rohde &amp; Schwarz SMR40 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Rohde &amp; Schwarz SMR40. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Rohde &amp; Schwarz SMR40 faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Rohde &amp; Schwarz SMR40 price

This Rohde &amp; Schwarz SMR40 is listed at $6,400. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Rohde &amp; Schwarz SMR40 calibration cost

NIST-traceable calibration for the Rohde &amp; Schwarz SMR40 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents